Teledyne 7300A 2.3Electrical Connections Rear Panel, Pipe Connections, Sample Delivery System

Models: 7300A

1 85
Download 85 pages 46.59 Kb
Page 15
Image 15
2.2.4 Pipe Connections

Infrared Gas Analyzer

Installation 2

 

 

 

 

 

 

 

 

 

2.2.4 Pipe Connections

Refer to Appendix Piping Drawings for information about pipe connec- tions. On special systems, consult the text in the manual that describes your particular sample system in detail.

2.2.5 Sample Delivery System

The sample delivery system should be designed to operate reliably and must be of large enough capacity to avoid flow stops. A pump is required only if there is insufficient pressure to reliably supply the sample to the system equipment panel. Do not complicate the delivery system by adding a pump unless it is absolutely necessary. If a pump is required, select a type that can handle the sample (corrosion), as well as meet the area classification and Environmental conditions.

2.2.6 Venting the System

In gas analysis systems, the system vent manifold or bypass/sample vents must terminate in a safe area as the sample may be poisonous, corrosive or flammable.

2.3Electrical Connections (Rear Panel)

Figure 3-3 shows the Model 7300A rear panel. There are connectors for power, digital communications, and both digital and analog concentration output.

For safe connections, no uninsulated wiring should be able to come in contact with fingers, tools or clothing during normal operation.

CAUTION: Use Shielded Cables. Also, use plugs that provide excellent EMI/RFI protection. The plug case must be connected to the cable shield, and it must be tightly fastened to the analyzer with its fastening screws. Ultimately, it is the installer who ensures that the connections provide adequate EMI/RFI shielding.

Teledyne Analytical Instruments

2-3

Page 15
Image 15
Teledyne 7300A manual 2.3Electrical Connections Rear Panel, Pipe Connections, Sample Delivery System, Venting the System