Infrared Gas Analyzer | Operation 4 | |
|
|
|
|
|
|
|
|
|
From the MAIN MENU Screen—
1.Use UP/DOWN to select ALGORITHM , and Enter.
2.Select and Enter SETUP.
3.Select MANUAL from the Calibration Mode Select screen.
Dpt | INPUT | OUTPUT |
1 | Ø.ØØ | Ø.ØØ |
The data entry screen resembles the verify screen, but the gas values can be modified and the
Use the UP/DOWN key to set the INPUT value for the lowest concentra- tion into the first point. ENTER to move to OUTPUT field. Use the UP/ DOWN key to set the OUTPUT value, the lowest concentration into the first point. ENTER to accept the first setting.
After each point is entered, the
Dpt | INPUT | OUTPUT |
0 | Ø.ØØ | Ø.ØØ |
Repeat the above procedure for each of the data points you are setting (up to nine points:
When you are done, ESCAPE. The message, Completed. Wait for calcula- tion, appears briefly, and then the main System screen returns.
To end the session, send:
st<enter> st<enter>
to the analyzer from the computer.
4.8.2.3 Auto Mode Linearization
To linearize in the Auto Mode, you must have on hand a separate calibra- tion gas for each of the data points you are going use in your linearization. First, the analyzer is zeroed and spanned as usual. Then, each special calibration gas, for each of the intermediate calibration points, is flowed, in turn, through the sensor. As each gas flows, the differential value for that intermediate calibration point is entered from the front panel of the analyzer.