Teledyne 7300A Precautions, Infrared Gas Analyzer, Operation, Teledyne Analytical Instruments

Models: 7300A

1 85
Download 85 pages 46.59 Kb
Page 61
Image 61
4.6.3 Precautions

Infrared Gas Analyzer

Operation 4

 

 

 

 

 

 

 

 

 

range readings are accurate UNLESS the application uses linearization over the selected range.

The concentration ranges can be redefined using the Range function Manual screen, and the application gases can be redefined using the APPLICA- TION function, if they are not already defined as necessary.

CAUTION: Redefining applications or ranges might require re- linearization and/or recalibration.

Tosetupautomaticranging:

Select range on the MAIN MENU, and ENTER to start the Range func-

tion.

Select range mode : AUTO

Note: If all three ranges are currently defined for different applica- tion gases, then the above screen does not display (because mode must be manual).

If above screen displays, use the UP/DOWN key to Select AUTO, and Enter.

Press Escape to return to the previous Analyze Function.

4.6.3 Precautions

The Model 7300A allows a great deal of flexibility in choosing ranges for automatic range switching. However, there are some pitfalls that are to be avoided.

Ranges that work well together are:

Ranges that have the same lower limits but upper limits that differ by approximately an order of magnitude

Ranges whose upper limits coincide with the lower limits of the next higher range

Ranges where there is a gap between the upper limit of the range and the lower limit of the next higher range.

Range schemes that are to be avoided include:

Teledyne Analytical Instruments

4-25

Page 61
Image 61
Teledyne 7300A manual Precautions, Infrared Gas Analyzer, Operation, Teledyne Analytical Instruments