Teledyne 7300A manual Trouble Shooting Chart, Infrared Gas Analyzer, Maintenance

Models: 7300A

1 85
Download 85 pages 46.59 Kb
Page 77
Image 77
5.8 TROUBLE SHOOTING CHART

Infrared Gas Analyzer

Maintenance 5

 

 

 

 

 

 

 

 

 

5.8 TROUBLE SHOOTING CHART

SYMPTOM

CORRECTIVE ACTION

A. Unit does not turn ON when actuating Power Switch (Motor does not turn).

B. No Zero Control (with unit on Zero gas). Out- put is either dead or fully saturated.

(a)Check AC voltage at inlet for correct voltage. Reset breaker or switch to supply power to receptacle.

(b)Check power to terminal strip located inside cabinet. CAUTION: Terminal strip contains 90-220volt line voltage which can be hazar- dous. Check for loose connec- tion to one of the terminal board lugs; replace as required.

(a)Check that Power is ON.

(b)Check that power supply voltages are correct (+15VDC, Ä15VDC).

(c)Check output of TP6 on preamp board with a DC voltmeter (should be less than + or Ä 3VDC). If output can be zeroed, check volt- ages at TPl and TP2 on Linearizer board (should also be zero). Then check adjustments on E/I board.

(d)Check that infrared source is operating. If no light, check source element resistance and connections.

(e) Try to execute a manual zero.

Teledyne Analytical Instruments

5-7

Page 77
Image 77
Teledyne 7300A manual Trouble Shooting Chart, Infrared Gas Analyzer, Maintenance, Teledyne Analytical Instruments