Teledyne 7300A Infrared Gas Analyzer, Installation, Pin Contact, 2.3.3.3Digital Remote Cal Inputs

Models: 7300A

1 85
Download 85 pages 46.59 Kb
Page 19
Image 19
Pin Contact

Infrared Gas Analyzer

Installation 2

 

 

 

 

 

 

 

 

 

System Alarm:

Actuates when DC power supplied to circuits is

 

unacceptable in one or more parameters. Permanently

 

configured as failsafe and latching. Cannot be de-

 

feated. Actuates if self test fails.

 

(Reset by pressing

button to remove power. Then

 

press

again and any other button EXCEPT

System to resume.

Further detail can be found in chapter 4, section 4-5.

Table 2-5: Alarm Relay Contact Pins

Pin Contact

45Threshold Alarm 1, normally closed contact

28Threshold Alarm 1, moving contact

46Threshold Alarm 1, normally open contact

42Threshold Alarm 2, normally closed contact

44Threshold Alarm 2, moving contact

43Threshold Alarm 2, normally open contact

36System Alarm, normally closed contact

20System Alarm, moving contact

37System Alarm, normally open contact

2.3.3.3Digital Remote Cal Inputs

Accept 0 V (off) or 24 V dc (on) inputs for remote control of calibra- tion. (See Remote Calibration Protocol below.) See Table 2-5 for pin connections.

Zero:

Floating input. A 5 to 24 V pulse input across the + and –

 

pins puts the analyzer into the Zero mode. Either side may be

 

grounded at the source of the signal. A synchronous signal

 

must open and close the gas control valves appropriately.

Span:

Floating input. A 5 to 24 V pulse input across the + and –

 

pins puts the analyzer into the Span mode. Either side may

 

be grounded at the source of the signal. A synchronous signal

 

must open and close the gas control valves appropriately.

Cal Contact: This relay contact is closed while analyzer is spanning and/or zeroing. (See Remote Calibration Protocol below.)

Teledyne Analytical Instruments

2-7

Page 19
Image 19
Teledyne 7300A manual Infrared Gas Analyzer, Installation, Pin Contact, 2.3.3.3Digital Remote Cal Inputs