VXI Technology, Inc.

 

 

OUTPut:POLarity:EXTernal:INTerrupt

 

 

 

 

 

 

Purpose

Sets the interrupt output polarity on the front panel

 

 

 

 

 

 

Type

Setting

 

 

 

 

 

 

Command Syntax

OUTPut:POLarity:EXTernal:INTerrupt <polarity>

 

 

 

 

 

 

Command Parameters

<polarity> = NORMal INVert

 

 

 

 

 

 

*RST Value

NORMal

 

 

 

 

 

 

Query Syntax

OUTPut:POLarity:EXTernal:INTerrupt?

 

 

 

 

 

 

Query Parameters

N/A

 

 

 

 

 

 

Query Response

ASCII string = NORM ⏐ INV

 

 

 

 

 

 

Description

The OUTput:POLarity:EXTernal:INTerrupt command sets the polarity of the front

 

 

 

panel interrupt output. When the polarity is set for normal, the output will be high

 

 

 

when there is an interrupt event. When set for invert, the output will be low when there

 

 

 

is an interrupt event.

 

 

 

 

 

 

 

 

 

 

Examples

Command / Query

Response (Description)

 

 

OUTP:POL:EXT:INT NORM

(Sets the front panel interrupt output polarity to normal)

 

 

OUTP:POL:EXT:INT?

NORM (Returns the set value for the front panel

 

 

 

interrupt output polarity)

 

 

 

 

 

Related Commands

None

 

76

VM4016 Command Dictionary

Page 76
Image 76
VXI VM4016 user manual OUTPutPOLarityEXTernalINTerrupt