www.vxitech.com

 

 

OUTPut:POLarity:EXTernal:LATChed

 

 

 

 

 

 

Purpose

Sets the latched interrupt output polarity on the front panel

 

 

 

 

 

 

Type

Setting

 

 

 

 

 

 

Command Syntax

OUTPut:POLarity:EXTernal:LATChed <polarity>

 

 

 

 

 

 

Command Parameters

<polarity> = NORMal INVert

 

 

 

 

 

 

*RST Value

NORMal

 

 

 

 

 

 

Query Syntax

OUTPut:POLarity:EXTernal:LATChed?

 

 

 

 

 

 

Query Parameters

N/A

 

 

 

 

 

 

Query Response

ASCII string = NORM ⏐ INV

 

 

 

 

 

 

Description

The OUTput:POLarity:EXTernal:LATChed command sets the polarity of the front

 

 

 

panel latched interrupt output. When the polarity is set for normal, the output will be

 

 

 

high when there is an interrupt event. When set for invert, the output will be low

 

 

 

when there is an interrupt event.

 

 

 

 

 

 

 

 

 

 

Examples

Command / Query

Response (Description)

 

 

OUTP:POL:EXT:LATC INV

(Sets the polarity of the front panel latched interrupt

 

 

 

output to inverted)

 

 

OUTP:POL:EXT:LATC?

INV (Returns the value for the front panel latched

 

 

 

interrupt output)

 

 

 

 

 

Related Commands

None

 