VXI Technology, Inc.

 

 

INPut:POLarity

 

 

 

 

 

 

Purpose

Sets the input polarity for one or more channels

 

 

 

 

 

 

Type

Setting

 

 

 

 

 

 

Command Syntax

INPut:POLarity <polarity>,<channel_list>

 

 

 

 

 

 

Command Parameters

<polarity> = NORMal INVert

 

 

 

<channel_list> = standard channel list syntax supporting channels 1 to 16

 

 

 

 

 

 

*RST Value

NORMal for all channels

 

 

 

 

 

 

Query Syntax

INPut:POLarity? <channel>

 

 

 

 

 

 

Query Parameters

<channel> = 1 to 16

 

 

 

 

 

 

Query Response

ASCII string = NORM INV

 

 

 

 

 

 

Description

The INPut:POLarity command selects the input polarity for one or more channels.

 

 

 

When a channel is programmed for normal polarity, an interrupt will occur when the

 

 

 

input voltage is greater than the programmed input offset for the channel. The invert

 

 

 

polarity will cause an interrupt when the input voltage is less than the programmed

 

 

 

input offset for the channel.

 

 

 

 

 

 

 

 

 

 

Examples

Command / Query

Response (Description)

 

 

INP:POL INV,(@5:12)

(Inverts the input polarity for Channels 5 - 12)

 

 

INP:POL? 6

INV (Indicates the polarity for Channel 6 is inverted)

 

 

 

 

 

Related Commands

INPut:OFFset

 

 

 

INPut:RANGe

 

74

VM4016 Command Dictionary

Page 74
Image 74
VXI VM4016 user manual INPutPOLarity