VXI Technology, Inc.

 

 

INPut:DEBounce

 

 

 

 

 

 

Purpose

Sets the debounce time

 

 

 

 

 

 

Type

Setting

 

 

 

 

 

 

Command Syntax

INPut:DEBounce <value>

 

 

 

 

 

 

Command Parameters

<value> = 9.6 µs to 0.6291456 s

 

 

 

 

 

 

*RST Value

19.2 µs

 

 

 

 

 

 

Query Syntax

INPut:DEBounce?

 

 

 

 

 

 

Query Parameters

N/A

 

 

 

 

 

 

Query Response

Numeric ASCII value from 0.0000096 to 0.6291456

 

 

 

 

 

 

Description

The INPut:DEBounce command sets the time period for the digital debounce circuitry.

 

 

 

By programming a debounce time of 1 ms, an input must exceed its threshold level for

 

 

 

a period of 1 ms before it is recognized as a valid input. The debounce resolution is

 

 

 

9.6 µs. The debounce time set is applied to all channels.

 

 

 

 

 

 

 

 

Examples

Command / Query

Response (Description)

 

INP:DEB 9.6e-6

(Sets a digital debounce time of 9.6 µs)

 

INP:DEB

0.0000096 (Indicates that the debounce time is set to

 

 

9.6 µs)

 

 

 

Related Commands

None

 

70

VM4016 Command Dictionary

Page 70
Image 70