Status Reporting

This section describes the complete status model of the instrument. Note that some registers are specific to the GPIB section of the instrument and are of limited use in an RS232 environment.

Standard Event Status and Standard Event Status Enable Registers

These two registers are implemented as required by the IEEE Std. 488.2.

Any bits set in the Standard Event Status Register which correspond to bits set in the Standard Event Status Enable Register will cause the ESB bit to be set in the Status Byte Register.

The Standard Event Status Register is read and cleared by the *ESR? command. The Standard Event Status Enable register is set by the *ESE <nrf> command and read by the *ESE? command.

Bit 7 - Power On. Set when power is first applied to the instrument. Bit 6 - Not used.

Bit 5 - Command Error. Set when a syntax type error is detected in a command from the bus.

The parser is reset and parsing continues at the next byte in the input stream.

Bit 4 - Execution Error. Set when an error is encountered while attempting to execute a completely parsed command. The appropriate error number will be reported in the Execution Error Register.

1- 99 Indicates a hardware error has been encountered.

116A recall of set up data has been requested but the store specified does not contain any data.

117A recall of set up data has been requested but the store specified contains corrupted data. This indicates either a hardware fault or a temporary data corruption which can be corrected by writing data to the store again.

120The numerical value sent with the command was too big or too small. Includes negative numbers where only positive numbers are accepted.

123A recall/store of set up data has been requested from/to an illegal store number.

124A range change has been requested but the current psu settings make it illegal – see manual operation instructions for details.

Bit 3 - Verify Timeout Error. Set when a parameter is set with 'verify' specified and the value is not reached within 5 secs, e.g. output voltage is slowed by a large capacitor on the output.

Bit 2 - Query Error. Set when a query error occurs. The appropriate error number will be reported in the Query Error Register as listed below.

1.Interrupted error

2.Deadlock error

3.Unterminated error

Bit 1 - Not used.

Bit 0 - Operation Complete. Set in response to the *OPC command.

Limit Event Status Registers and Limit Event Status Enable Registers

Two pairs of registers are implemented as an addition to the IEEE Std.488.2. Each pair consists of a Limit Event Status Register and an accompanying Limit Status Event Enable Register. Limit Event Status Register 1 (LSR1) and Limit Event Status Enable Register 1 (LSE1) apply to output 1. Limit Event Status Register 2 (LSR2) and Limit Event Status Enable Register 2 (LSE2) apply to output 2 and the Auxiliary output. Their purpose is to inform the controller of entry to and/or exit from current or voltage limit conditions by storing a history of protection trip conditions since the last read.

Any bits set in a Limit Event Status Register which correspond to bits set in the accompanying Limit Event Status Enable Register will cause the LIM1 or LIM2 bit to be set in the Status Byte Register.

33

Page 34
Image 34
Xantrex Technology XDL 35-5TP manual Status Reporting

XDL 35-5T, XDL 35-5TP specifications

Xantrex Technology has established itself as a leader in the power conversion industry, and their XDL series, specifically the XDL 35-5TP and XDL 35-5T, exemplifies their commitment to innovation and reliability in power management solutions. Designed for a variety of applications, these models offer high performance and efficiency, catering to both professional and recreational users.

The XDL 35-5TP is recognized for its robust design and versatility, suitable for various applications, including battery charging, powering electronic devices, and supporting other power equipment. It features a maximum output of 35 amps and can deliver a continuous power supply sufficient for multiple devices. The integrated temperature monitoring ensures optimal performance and longevity, while the built-in safety features protect against overloads and short circuits.

One of the standout features of the XDL 35-5TP is its advanced microprocessor control, which allows for smart charging profiles. This technology not only increases efficiency but also extends battery life by managing the charging cycle meticulously. Users can take advantage of adjustable charging voltage settings, making the XDL 35-5TP compatible with various battery types, including lead-acid and lithium-ion.

The XDL 35-5T, on the other hand, while sharing many of the core features of the TP model, focuses more on size and weight optimization without compromising power delivery. This makes it an excellent choice for mobile users and those with limited installation space. Both models boast a durable design, suitable for harsh environmental conditions, ensuring reliability whether used in marine, automotive, or off-grid applications.

In terms of connectivity, the XDL series includes multiple output options, allowing users to connect various devices seamlessly. The front-mounted display panel gives real-time feedback on battery status and system performance, enabling informed management of power resources. Additionally, both models are engineered to minimize electromagnetic interference, ensuring operation without disrupting other electronic systems.

Overall, the Xantrex XDL 35-5TP and XDL 35-5T represent cutting-edge solutions in power technology, combining innovative features with practical usability. Their high efficiency, safety protocols, and flexibility in applications make them an excellent choice for users seeking reliable power solutions in diverse settings. Whether for recreational vehicles, boats, or emergency backup, the XDL series stands out as a testament to Xantrex's commitment to quality and performance in power conversion technology.