RANGE LOCK
This instrument has a Range Lock feature that displays readings to the best resolution, shown by A/R on the display. This can be turned off by pressing the A/R key. The display will then use the same display resolution for the entire pressure range. This is useful to stop the display switching between resolutions where there are fluctuations in pressure. To cancel this mode, press the A/R key again.
ADVANCED FUNCTIONS
The following additional functions apply to models 2080P, 2081P, 2082P, 2083P, 2084P, 2085P
LOGGING
These models have functions that enable the user to store and retrieve up to 250 readings and output such to a PC or Epson compatible printer via
offer DigiLog, a customised Windows™ software package.
Note: Some functions may be inhibited while information is stored in either Auto, Manual or Max/Min modes. This is to prevent inadvertent parameter mixing whilst in any of the above modes.
To
SETTING INTERNAL CLOCK
Press decimal point (•) key and the and NE symbols will appear on display with the current year and month settings (YYMM).
Key in new year and month if desired and press key to accept. Display will show current date of month setting (DD). Key in new value if desired and press key to accept.
Display will show current hours and minutes setting (24 hour clock). Key in new values if desired and press key to accept. Instrument will then return to normal mode.
LOGGING ON DEMAND
This function allows readings to be stored as and when required. A maximum of 50 readings can be stored this way.
When ready to store a reading press key. , NE symbols and 0 will appear on the display. It is possible to enter up to a four digit reference number including decimal point. Press key to enter and store reading, time, date and reference number.
Instrument will return to standard mode, however symbol will remain on display until data is erased.
PRESET INTERVAL LOGGING
This function allows readings to be taken automatically at a preset time interval.
NOTE: If data is already stored, as signified by the or symbols appearing on display, this function will not operate.
Press key. NE symbol and 0000 will appear on the display. Enter required time interval in hh:mm (ie, in order to take a reading every 1 minute, key in 0001) and press key. Current reading will
return to display and symbol will remain on display. Instrument will continue logging until a maximum of 250 readings are taken or data is