Signal Equalizer Settings for Non Recommended Cable or Unknown Cable Lengths, Continued
View Measurements
Signals are viewed with an oscilloscope.
Scope characteristics: 10 MHz bandwidth capable of measurement sen- sitivity of 0.2 v/div to 1.0 V/div. A typical
Option 1. Connection Point
Option 2. Connection Point
Connect the two scope probes to the positive and negative terminals of the HUB Communications board, not the Equalizer board, of the channel to be measured. Alternatively, a connection can be made to the leads of the fuses for the selected channel.
•Set time base to ~1 or 2 millisecond per division.
•Observe two sequential signal envelopes of
The first signal envelope is the poll from the HUB to the slave device, and the second is the response back from the slave device. The reading from the poll should be ~5 volts
If a large amount of common mode line frequency interference is present the above choice may be unsuitable. In this case, Option 2. is recom- mended.
This option observes the signal across the termination resistor for the selected port.
Using this option, connect the two scope probes across an unused im- pedance jumper location. For instance, if the jumper is in the
Both channels should be
Using this method there is only one reading, the response message, which is seen. Responses may be from ~5 volts
38 • Installation |