Teledyne M5000B operating instructions Select More and press the Enter Key

Page 33

Photometric Analyzer

Operation /Control Unit 3

 

 

 

 

 

 

 

 

 

Note: Before setting up an AUTO-CAL, be sure you understand the Zero and Span functions as described in section 4.4, and follow the precautions given there.

Note: If you require highly accurate AUTO-CAL timing, use external AUTO-CAL control where possible. The internal clock in the Model 5000BF is accurate to 2-3 %. Accordingly, internally scheduled calibrations can vary 2-3 % per day.

To setup an Auto–Cal cycle:

Choose System from the Function buttons. TheVFD will display five subfunctions.

DIG_FILT SELF—TEST

PWD LOGOUT MORE

Select MORE and press the Enter Key

AUTOCAL

FILSOL

HOLD

CAL-HOLD-TIMER

MORE

 

Use < > arrows to blink AUTO—CAL, and press Enter. A new screen for ZERO/SPAN set appears.

ZERO in Ød Øh off

SPAN in Ød Øh off

Press < > arrows to blink ZERO (or SPAN), then press Enter again. (You won’t be able to set OFF to ON if a zero interval is entered.) A Span Every ... (or Zero Every ...) screen appears.

Zero schedule: OFF

Day: Ød Hour: Øh

Use DÑ arrows to set an interval value, then use < > arrows to move to the start-time value. Use DÑ arrows to set a start-time value.

To turn ON the SPAN and/or ZERO cycles (to activate AUTO–CAL): Press System again, choose AUTO–CAL,and press Enter again. When the ZERO/SPAN values screen appears, use the < > arrows to blink the ZERO (or SPAN) and press Enter to go to the next screen. Use < > to select OFF/ ON field. Use DÑ arrows to set the OFF/ON field to ON. You can now turn these fields ON because there is a nonzero span interval defined.

If instrument is turned off, the next time the instrument is powered, the instrument will automatically perform a calibration cycle after 3 minutes of entering the sample mode if AUTOCAL functions were on prior to shut down.

Teledyne Analytical Instruments

Part I 3-5

Image 33
Contents Teledyne Analytical Instruments Photometric AnalyzerCopyright 1999 Teledyne Analytical Instruments Model 5000BIii Table of ContentsModel 5000B Teledyne Analytical Instruments Part I Control Unit Part IInstallation Ii PartOperation Maintenance Model 5000B Photometric Analyzer Iv PartOverview Typical ApplicationsPhotometric Analyzer Part I Control Unit Introduction Model 5000B Other NIR Absorbers that can be measured consult factory Introduction Model 5000B Main Features of the Analyzer 1 UP/DOWN Switch ESCAPE/ENTER SwitchOperator Interface Model 5000B Controls, Indicators, and Connectors Control Section Interface Panel Model 5000B Interface Panel Introduction Model 5000B Unpacking the Control/Analysis Unit InstallationPart 115VAC Primary Input Power ConnectionsAnalog Output Connections ExamplesRange Voltage Current mA Alarm RelaysTypes of Relay Contacts The connectors are Part I RS-232 Sig RS-232 Pin Purpose RS-232 SignalsTesting the System Command DescriptionParameter Setting Part Photometric Analyzer Operation /Control Unit Operation Operation/Control Unit Hierarchy of System Functions and Subfunctions System Function Setting up an AUTO-CALSelect More and press the Enter Key Entering the Password PasswordProtectionInstalling or Changing the Password Logging Out System Self-Diagnostic Test CheckingLinearitywithALGORITHM Model ScreenSetting 90% Response time seconds Digital Filter SetupFilter or Solenoid Setup Calibration/Hold Timer Setup Hold/Track SetupAnalog 4 to 20 mA Output Calibration Calbrt hold 3 min Sample hold 1 minModel Use UP/DOWN arrow to Adjust 4 ma 250Auto Mode Zeroing Show NegativeZero Cal Manual Mode Zeroing Sccm Cell FailureAuto Mode Spanning Manual Mode SpanningSpan Cal Offset Function How to access the offset functionZero off 0.0 ppm ENT to begin Zero Select zero mode Auto Select zero mode MANAlarms Function AL1 1ØØØ ppm HI DftN FsN LtchN Manual Select/Define Range Screen Range Select FunctionAuto Screen Precautions 2illustratestheseschemesgraphically Analyze Function Set Range Screen Algorithm Application More Curve Algorithm Screen Auto Mode Linearization Part I Control Unit/Analysis Unit Maintenance Fuse ReplacementMaintenance Model 5000B Photometric Analyzer System Self Diagnostic TestPower AnalogCell detector PreampPart Part II Analysis Unit Part IIRoutine Maintenance Part II Analysis Unit Appendix Iv Part Operations/Analysis Unit Explosion-ProofVersionControlModule Control FunctionsStart-p Control SettingsPreliminaryInspection Analyzers not equipped with Auto ZeroPre-Start-upElectricalCheckout Power On ObservationsTeledyneAnalyticalInstruments4-5 Procedure CalibrationStandardizationFluids TeledyneAnalyticalInstruments4-7 Operations/Analysis Unit TeledyneAnalyticalInstruments4-9 Operations/Analysis Unit TeledyneAnalyticalInstruments4-11 Operations/Analysis Unit TeledyneAnalyticalInstruments4-13 Ppm simulated water using Sample Span ZeroAttachment 5000B/5020 NIR Analyzer Operations/Analysis Unit Operational Theory Contd Source Module Sample cell OpticalSystemAnalyzerSystem-BlockDiagram Power Module Block Diagram Detectorcompartment Part II Maintenance/Analysis Unit Routine MaintenanceSystem Visual Check and Response Procedure Automatic operation and routine operational dutiesDaily Suggested Preventive Maintenance ScheduleEquipment Required Oscilloscope Display of the I to E Converter Output Part II Maintenance/Analysis Unit Inverting Amplifier Setup of the Logarithmic AmplifierInterface Board Terminals Strip Battery-Powered Oscilloscope Synchronization PointPart Part II Maintenance/Analysis Unit Part Specifications Operating Temperature 0-50oCPhotometric Analyzer, 5000B Appendix Appendix Models 5000B Other Features QtyP/NDescription Recommended 2-Year Spare Parts ListDrawing List, 5000B Appendix Models 5000B Teledyne Analytical Instruments