Photometric Analyzer | Operation /Control Unit 3 | |
|
|
|
|
|
|
|
|
|
AUTOCAL FILSOLL HOLD
Select MORE and press ENTER one more time
ALGORITHM APPLICATION
MORE OUTPUT: 4MA
Now you will be able to select the APPLICATION and ALGO- RITHM
3.8.1The Set Range Screen
The Set Range screen allows reprogramming of the three analysis ranges and the calibration range (background fluid, low end of range, high end of range, and % or ppm units). Original programming is usually done at the factory according to the customer’s application. It must be done through the
Note: It is important to distinguish between this System programming subfunction and the Range button function, which is an operator control. The Set Range Screen of the System function allows the user to DEFINE the upper and lower limits of a range AND the application of the range. The Range button function only allows the user to select or define the limits, or to select the application, but not to define the application.
Normally the Model 5000BF is factory set to default to manual range selection, unless it is ordered as a
In the autoranging mode, the microprocessor automatically responds to concentration changes by switching ranges for optimum readout sensitivity. If the upper limit of the operating range is reached, the instrument automati- cally shifts to the next higher range. If the concentration falls to below 85% of full scale of the next lower range, the instrument switches to the lower range. A corresponding shift in the DC concentration output, and in the range ID outputs, will be noticed.
The autoranging feature can be overridden so that analog output stays on a fixed range regardless of the contaminant concentration detected. If the concentration exceeds the upper limit of the range, the DC output will saturate at 1 V dc (20 mA at the current output).
However, the digital readout and the
Teledyne Analytical Instruments | Part I |