Teledyne M5000B operating instructions Model Screen, CheckingLinearitywithALGORITHM

Page 38

3 Operation/Control Unit

Model 5000B

 

 

 

 

 

 

 

 

 

3.3.5 The Model Screen

Move the < > arrow key to MORE and press Enter. With MODEL blinking, press Enter. The screen displays the manufacturer, model, and softwareversioninformation.

3.3.6CheckingLinearitywithALGORITHM

From the System Function screen, select ALGORITHM, and press Enter.

sel rng to set algo: —> Ø1 Ø2 Ø3 <—

Use the < > keys to select the range: 01, 02, or 03. Then press Enter.

Fluid Use: SO2

Range: Ø — 10%

Press Enter again.

Algorithm setup:

VERIFY SET UP

Select and Enter VERIFY to check whether the linearization has been accomplished satisfactorily.

Dpt INPUT OUTPUT Ø Ø.ØØ Ø.ØØ

The leftmost digit (under Dpt) is the number of the data point being monitored. Use the DÑ keys to select the successive points.

The INPUT value is the input to the linearizer. It is the simulated output of the analyzer. You do not need to actually flow fluid.

The OUTPUT value is the output of the linearizer. It should be the ACTUAL concentration of the span fluid being simulated.

If the OUTPUT value shown is not correct, the linearization must be corrected. Press ESCAPE to return to the previous screen. Select and Enter SET UP to Calibration Mode screen. (set-up will not work without a PC being connected to the analyzer)

Select algorithm mode : AUTO

There are two ways to linearize: AUTO and MANUAL: The auto mode requires as many calibration fluides as there will be correction points along the curve. The user decides on the number of points, based on the preci- sionrequired.

3-10 Part I

Teledyne Analytical Instruments

Image 38
Contents Photometric Analyzer Teledyne Analytical InstrumentsModel 5000B Copyright 1999 Teledyne Analytical InstrumentsTable of Contents IiiModel 5000B Teledyne Analytical Instruments Part I Part I Control UnitOperation InstallationIi Part Maintenance Iv Part Model 5000B Photometric AnalyzerPhotometric Analyzer Part I Control Unit OverviewTypical Applications Introduction Model 5000B Other NIR Absorbers that can be measured consult factory Introduction Model 5000B Main Features of the Analyzer Operator Interface 1 UP/DOWN SwitchESCAPE/ENTER Switch Model 5000B Controls, Indicators, and Connectors Control Section Interface Panel Model 5000B Interface Panel Introduction Model 5000B Installation Unpacking the Control/Analysis UnitPart Primary Input Power Connections 115VACExamples Analog Output ConnectionsAlarm Relays Range Voltage Current mATypes of Relay Contacts The connectors are Part I RS-232 Signals RS-232 Sig RS-232 Pin PurposeParameter Setting Testing the SystemCommand Description Part Operation Photometric Analyzer Operation /Control UnitOperation/Control Unit Hierarchy of System Functions and Subfunctions Setting up an AUTO-CAL System FunctionSelect More and press the Enter Key PasswordProtection Entering the Password Installing or Changing the Password Logging Out System Self-Diagnostic Test Model Screen CheckingLinearitywithALGORITHMDigital Filter Setup Setting 90% Response time secondsFilter or Solenoid Setup Hold/Track Setup Calibration/Hold Timer SetupCalbrt hold 3 min Sample hold 1 min Analog 4 to 20 mA Output CalibrationUse UP/DOWN arrow to Adjust 4 ma 250 ModelZero Cal Auto Mode ZeroingShow Negative Manual Mode Zeroing Cell Failure SccmSpan Cal Auto Mode SpanningManual Mode Spanning How to access the offset function Offset FunctionSelect zero mode Auto Select zero mode MAN Zero off 0.0 ppm ENT to begin ZeroAlarms Function AL1 1ØØØ ppm HI DftN FsN LtchN Range Select Function Manual Select/Define Range ScreenAuto Screen Precautions 2illustratestheseschemesgraphically Analyze Function Set Range Screen Algorithm Application More Curve Algorithm Screen Auto Mode Linearization Fuse Replacement Part I Control Unit/Analysis Unit MaintenancePower System Self Diagnostic TestMaintenance Model 5000B Photometric Analyzer AnalogPreamp Cell detectorPart Part II Part II Analysis UnitRoutine Maintenance Part II Analysis Unit Appendix Iv Part Operations/Analysis Unit Control Functions Explosion-ProofVersionControlModulePreliminaryInspection Control SettingsStart-p Analyzers not equipped with Auto ZeroPower On Observations Pre-Start-upElectricalCheckoutTeledyneAnalyticalInstruments4-5 StandardizationFluids ProcedureCalibration TeledyneAnalyticalInstruments4-7 Operations/Analysis Unit TeledyneAnalyticalInstruments4-9 Operations/Analysis Unit TeledyneAnalyticalInstruments4-11 Operations/Analysis Unit TeledyneAnalyticalInstruments4-13 Ppm simulated water using Span Zero SampleAttachment 5000B/5020 NIR Analyzer Operations/Analysis Unit Operational Theory Contd Source Module OpticalSystem Sample cellAnalyzerSystem-BlockDiagram Power Module Block Diagram Detectorcompartment System Visual Check and Response Procedure Routine MaintenancePart II Maintenance/Analysis Unit Automatic operation and routine operational dutiesSuggested Preventive Maintenance Schedule DailyEquipment Required Oscilloscope Display of the I to E Converter Output Part II Maintenance/Analysis Unit Setup of the Logarithmic Amplifier Inverting AmplifierBattery-Powered Oscilloscope Synchronization Point Interface Board Terminals StripPart Part II Maintenance/Analysis Unit Part Photometric Analyzer, 5000B Appendix SpecificationsOperating Temperature 0-50oC Appendix Models 5000B Other Features Recommended 2-Year Spare Parts List QtyP/NDescriptionDrawing List, 5000B Appendix Models 5000B Teledyne Analytical Instruments