3 Operation/Control Unit | Model 5000B | |
|
|
|
|
|
|
|
|
|
####.## % SO2 Zero adj:2048
The analyzer goes through
####.## % SO2 Slope=#.###
Generally, you have a good zero when Slope is less than 0.05 ppm/s for about 30 seconds.
Once zero settling completes, the information is stored in the analyzer’s memory, and the instrument automatically returns to the Analyze mode.
3.4.1.3 Cell Failure
Detector failure in the 5000BF is usually associated with inability to zero the instrument with a reasonable voltage differential between the refer- ence and measure voltages. If this should ever happen, the 5000BF system alarm trips, and the LCD displays a failure message.
Detector cannot be balanced
Check your zero fluid
Before optical balancing:
a.Check your zero fluid to make sure it is within specifications.
b.Check for leaks downstream from the Sample Cell, where con- tamination may be leaking into the system.
c.Check flowmeter to ensure that the flow is no more than 200
SCCM
d.Check temperature controller board.
e.Check fluid temperature.
f.Check the Sample Cell for dirty windows.
If none of the above, proceed to perform an optical balance as described in section 5.