Teledyne M5000B operating instructions Manual Mode Zeroing

Page 45

Photometric Analyzer

Operation /Control Unit 3

 

 

 

 

 

 

 

 

 

zero calibration is to be performed automatically or manually. Use the DÑ arrow keys to toggle between AUTO and MAN zero settling. Stop when AUTO appears, blinking, on the display.

Select zero mode: AUTO

Press Enter to begin zeroing.

####.## % SO2 Slope=#.### C—Zero

The beginning zero level is shown in the upper left corner of the dis- play. As the zero reading settles, the screen displays and updates information on Slope= in percent/second (unless the Slope starts within the acceptable zero range and does not need to settle further). The system first does a coarse zero, shown in the lower right corner of the screen as C—Zero, for 3 min, and then does a fine zero, and displays F—Zero, for 3 min.

Then, and whenever Slope is less than 0.01 for at least 3 min, instead of Slope you will see a countdown: 9 Left, 8 Left, and so fourth. These are software steps in the zeroing process that the system must complete, AFTER settling, before it can go back to Analyze. Software zero is indicated by S– Zero in the lower right corner.

####.## % SO2

4 Left=#.### S—Zero

The zeroing process will automatically conclude when the output is within the acceptable range for a good zero. Then the analyzer automatically returns to the Analyze mode.

3.4.1.2Manual Mode Zeroing

Press Zero to enter the Zero function. The screen that appears allows you to select between automatic or manual zero calibration. Use the DÑ keys to toggle between AUTO and MAN zero settling. Stop when MANU- AL appears, blinking, on the display.

Select zero mode: MANUAL

Press Enter to begin the zero calibration. After a few seconds the first of three zeroing screens appears. The number in the upper left hand corner is the first-stage zero offset. The microprocessor samples the output at a prede- termined rate.

Teledyne Analytical Instruments

Part I 3-17

Image 45
Contents Teledyne Analytical Instruments Photometric AnalyzerCopyright 1999 Teledyne Analytical Instruments Model 5000BIii Table of ContentsModel 5000B Teledyne Analytical Instruments Part I Control Unit Part IInstallation Ii PartOperation Maintenance Model 5000B Photometric Analyzer Iv PartOverview Typical ApplicationsPhotometric Analyzer Part I Control Unit Introduction Model 5000B Other NIR Absorbers that can be measured consult factory Introduction Model 5000B Main Features of the Analyzer 1 UP/DOWN Switch ESCAPE/ENTER SwitchOperator Interface Model 5000B Controls, Indicators, and Connectors Control Section Interface Panel Model 5000B Interface Panel Introduction Model 5000B Unpacking the Control/Analysis Unit InstallationPart 115VAC Primary Input Power ConnectionsAnalog Output Connections ExamplesRange Voltage Current mA Alarm RelaysTypes of Relay Contacts The connectors are Part I RS-232 Sig RS-232 Pin Purpose RS-232 SignalsTesting the System Command DescriptionParameter Setting Part Photometric Analyzer Operation /Control Unit OperationOperation/Control Unit Hierarchy of System Functions and Subfunctions System Function Setting up an AUTO-CALSelect More and press the Enter Key Entering the Password PasswordProtectionInstalling or Changing the Password Logging Out System Self-Diagnostic Test CheckingLinearitywithALGORITHM Model ScreenSetting 90% Response time seconds Digital Filter SetupFilter or Solenoid Setup Calibration/Hold Timer Setup Hold/Track Setup Analog 4 to 20 mA Output Calibration Calbrt hold 3 min Sample hold 1 minModel Use UP/DOWN arrow to Adjust 4 ma 250Auto Mode Zeroing Show NegativeZero Cal Manual Mode Zeroing Sccm Cell FailureAuto Mode Spanning Manual Mode SpanningSpan Cal Offset Function How to access the offset functionZero off 0.0 ppm ENT to begin Zero Select zero mode Auto Select zero mode MANAlarms Function AL1 1ØØØ ppm HI DftN FsN LtchN Manual Select/Define Range Screen Range Select FunctionAuto Screen Precautions 2illustratestheseschemesgraphically Analyze Function Set Range Screen Algorithm Application More Curve Algorithm Screen Auto Mode Linearization Part I Control Unit/Analysis Unit Maintenance Fuse ReplacementMaintenance Model 5000B Photometric Analyzer System Self Diagnostic TestPower AnalogCell detector PreampPart Part II Analysis Unit Part IIRoutine Maintenance Part II Analysis Unit Appendix Iv Part Operations/Analysis Unit Explosion-ProofVersionControlModule Control FunctionsStart-p Control SettingsPreliminaryInspection Analyzers not equipped with Auto ZeroPre-Start-upElectricalCheckout Power On ObservationsTeledyneAnalyticalInstruments4-5 Procedure CalibrationStandardizationFluids TeledyneAnalyticalInstruments4-7 Operations/Analysis Unit TeledyneAnalyticalInstruments4-9 Operations/Analysis Unit TeledyneAnalyticalInstruments4-11 Operations/Analysis Unit TeledyneAnalyticalInstruments4-13 Ppm simulated water using Sample Span ZeroAttachment 5000B/5020 NIR Analyzer Operations/Analysis Unit Operational Theory Contd Source Module Sample cell OpticalSystemAnalyzerSystem-BlockDiagram Power Module Block Diagram Detectorcompartment Part II Maintenance/Analysis Unit Routine MaintenanceSystem Visual Check and Response Procedure Automatic operation and routine operational dutiesDaily Suggested Preventive Maintenance ScheduleEquipment Required Oscilloscope Display of the I to E Converter Output Part II Maintenance/Analysis Unit Inverting Amplifier Setup of the Logarithmic AmplifierInterface Board Terminals Strip Battery-Powered Oscilloscope Synchronization PointPart Part II Maintenance/Analysis Unit Part Specifications Operating Temperature 0-50oCPhotometric Analyzer, 5000B Appendix Appendix Models 5000B Other Features QtyP/NDescription Recommended 2-Year Spare Parts ListDrawing List, 5000B Appendix Models 5000B Teledyne Analytical Instruments