Photometric Analyzer | Operation /Control Unit 3 | |
|
|
|
|
|
|
|
|
|
3.3.4System Self-Diagnostic Test
The Model 5000BF has a
Note: The sensor will always show failed unless Zero fluid is present in the sampling cell at the time of the
The self diagnostics are run automatically by the analyzer whenever the instrument is turned on, but the test can also be run by the operator at will. To initiate a self diagnostic test during operation:
Press the System button to start the System function.
DIG_FILT
PWD LOGOUT MORE
Use the < > arrow keys again to move the blinking to the
RUNNING DIAGNOSTIC
Testing Preamp — Cell
When the testing is complete, the results are displayed.
Power: OK Analog: OK
Cell: 2 Preamp: 3
The module is functioning properly if it is followed by OK. A number indicates a problem in a specific area of the instrument. Refer to Chapter 5 Maintenance and Troubleshooting for
Press Any Key
To Continue...
Then the analyzer returns to the initial System screen.