Teledyne M5000B operating instructions Part II Maintenance/Analysis Unit

Page 95

Part II: Maintenance/Analysis Unit

Maintenance 5

 

 

 

 

 

 

 

 

 

Theprocedureispurelymechanicalandconsistsofadjustingtheamountof light passing through either the measuringor reference filter, never both. Screens (wire mesh) of varying density are used for this operation and are part of the small tookkitaccompanyingtheinstrument.

1.Observe the oscilloscope and judge if optical balancing is needed. When the difference is less than 1 volt, balancing is not required. The tallest of the two peaks should be adjusted to 8 or 9 volts with the gain control R2 on the detector PC board. When this cannot be done because both peaks are too short or too long, search for screens mounted in the light path, usually located in a holder on the light pipe which interconnects the detector and sample module, and remove or add screens, as necessary.

2.When balancing is needed, identify the peaks as outlined under Section

3.For example, if the reference peak is the shorter one, stop the filter wheel with your hand and see if screens are located behind the reference filter. The reference filter is identified by the letter “R” engraved on the filter wheel.

4.If screens are found, remove them after taking the filter wheel off the shaft with the special Allen wrench supplied in the tool kit.

5.After removal of the screens and remounting the filter, mount the filter wheel back on the shaft. Position it correctly on the shaft by lining up the two paint marks on shaft and wheel.

6.Turn on the instrument and observe the balance on the oscillo- scope.

a.If the reference peak is now too tall, remove the filter wheel and add a screen of lesser density behind the reference filter. Repeat this procedure until the peaks are within 1 volt of each other.

b.If the measuring peak is equal to or within 1 volt of thereference peak, the system is optically balanced and ready for calibration.

c.If the peak is still too short, repeat the procedure, but thistime put a screen behind the measuring filter to shorten its peak.

Teledyne Analytical Instruments

Part II: 5-5

Image 95
Contents Teledyne Analytical Instruments Photometric AnalyzerCopyright 1999 Teledyne Analytical Instruments Model 5000BIii Table of ContentsModel 5000B Teledyne Analytical Instruments Part I Control Unit Part IOperation InstallationIi Part Maintenance Model 5000B Photometric Analyzer Iv PartPhotometric Analyzer Part I Control Unit OverviewTypical Applications Introduction Model 5000B Other NIR Absorbers that can be measured consult factory Introduction Model 5000B Main Features of the Analyzer Operator Interface 1 UP/DOWN SwitchESCAPE/ENTER Switch Model 5000B Controls, Indicators, and Connectors Control Section Interface Panel Model 5000B Interface Panel Introduction Model 5000B Unpacking the Control/Analysis Unit InstallationPart 115VAC Primary Input Power ConnectionsAnalog Output Connections ExamplesRange Voltage Current mA Alarm RelaysTypes of Relay Contacts The connectors are Part I RS-232 Sig RS-232 Pin Purpose RS-232 SignalsParameter Setting Testing the SystemCommand Description Part Photometric Analyzer Operation /Control Unit OperationOperation/Control Unit Hierarchy of System Functions and Subfunctions System Function Setting up an AUTO-CALSelect More and press the Enter Key Entering the Password PasswordProtectionInstalling or Changing the Password Logging Out System Self-Diagnostic Test CheckingLinearitywithALGORITHM Model ScreenSetting 90% Response time seconds Digital Filter SetupFilter or Solenoid Setup Calibration/Hold Timer Setup Hold/Track SetupAnalog 4 to 20 mA Output Calibration Calbrt hold 3 min Sample hold 1 minModel Use UP/DOWN arrow to Adjust 4 ma 250Zero Cal Auto Mode ZeroingShow Negative Manual Mode Zeroing Sccm Cell FailureSpan Cal Auto Mode SpanningManual Mode Spanning Offset Function How to access the offset functionZero off 0.0 ppm ENT to begin Zero Select zero mode Auto Select zero mode MANAlarms Function AL1 1ØØØ ppm HI DftN FsN LtchN Manual Select/Define Range Screen Range Select FunctionAuto Screen Precautions 2illustratestheseschemesgraphically Analyze Function Set Range Screen Algorithm Application More Curve Algorithm Screen Auto Mode Linearization Part I Control Unit/Analysis Unit Maintenance Fuse ReplacementAnalog System Self Diagnostic TestMaintenance Model 5000B Photometric Analyzer PowerCell detector PreampPart Part II Analysis Unit Part IIRoutine Maintenance Part II Analysis Unit Appendix Iv Part Operations/Analysis Unit Explosion-ProofVersionControlModule Control FunctionsAnalyzers not equipped with Auto Zero Control SettingsStart-p PreliminaryInspectionPre-Start-upElectricalCheckout Power On ObservationsTeledyneAnalyticalInstruments4-5 StandardizationFluids ProcedureCalibration TeledyneAnalyticalInstruments4-7 Operations/Analysis Unit TeledyneAnalyticalInstruments4-9 Operations/Analysis Unit TeledyneAnalyticalInstruments4-11 Operations/Analysis Unit TeledyneAnalyticalInstruments4-13 Ppm simulated water using Sample Span ZeroAttachment 5000B/5020 NIR Analyzer Operations/Analysis Unit Operational Theory Contd Source Module Sample cell OpticalSystemAnalyzerSystem-BlockDiagram Power Module Block Diagram Detectorcompartment Automatic operation and routine operational duties Routine MaintenancePart II Maintenance/Analysis Unit System Visual Check and Response ProcedureDaily Suggested Preventive Maintenance ScheduleEquipment Required Oscilloscope Display of the I to E Converter Output Part II Maintenance/Analysis Unit Inverting Amplifier Setup of the Logarithmic AmplifierInterface Board Terminals Strip Battery-Powered Oscilloscope Synchronization PointPart Part II Maintenance/Analysis Unit Part Photometric Analyzer, 5000B Appendix SpecificationsOperating Temperature 0-50oC Appendix Models 5000B Other Features QtyP/NDescription Recommended 2-Year Spare Parts ListDrawing List, 5000B Appendix Models 5000B Teledyne Analytical Instruments