Teledyne M5000B operating instructions Operations/Analysis Unit

Page 78

4 Operations/Analysis Unit

as the off-line method used to verify the process variables.

Below is an example of an on-line calibration when measuring 0-200ppm water in liquid TDI (assumes TDI is a toxic liquid at standard temperature and pressure conditions).

The first sample should be taken when the analyte (H20) concentration is known to be low (approximately O-.002 % water). The sample is quickly withdrawn and analyzed in the lab to determine the water content via an acceptable lab method (This may be Karl Fischer titration, gas chromatogra- phy, IR spectroscopy or volume extraction/distillation, etc.). At this time, the instrument zero is made to read exactly zero on the output of the analyzer (i.e., on the % or PPM of the panel/ digital meter, 4-20 ma current, or 0-1 VDC etc.) using the zero potentiometer of the analyzer or the digital or microprocessors zeroing calibration adjustment, Record the lab value, ana- lyzed corresponding to the zero set-point adjustment.

A second sample from the process must be grabbed after a steady state upscale reading has been reached. Hopefully, at this time, the customer can make a minor changes in the process to cause a slight upscale reading (1 0-30% of the full scale or more if possible) in order to obtain a steady-state reading at another concentration level. It this is not possible, then an appropriate time interval must occur when the process naturally changes to an upscale steady-state level concentra- tion. At this time the second calibration sample is grabbed and immediately analyzed. Record this value, This higher concen- tration value is subtracted from the first sample analyzed, Its difference value1 defines the slope of the calibration curve.

value1- This value will be used to determine the calibration slope of the instrument range.

Using the span adjustment of the analyzer, set the output reading to agree with this difference in concentration. Once the slope is set, adjust the zero of the instrument to agree with the true concentration to that of the analyzed steady-state sample presently flowing through the instrument cell. Now the analyzer is calibrated to the process fluid.

Zero calibration using secondary standard zero and span flag(s)

AFTER the analyzer has been calibrated on the process fluid, a simulation of this process on-line calibration can be made using internal synthetic

4–10 TeledyneAnalyticalInstruments

A Business Unit of Teledyne Electronic Technologies

Image 78
Contents Photometric Analyzer Teledyne Analytical InstrumentsModel 5000B Copyright 1999 Teledyne Analytical InstrumentsTable of Contents IiiModel 5000B Teledyne Analytical Instruments Part I Part I Control UnitInstallation Ii PartOperation Maintenance Iv Part Model 5000B Photometric AnalyzerOverview Typical ApplicationsPhotometric Analyzer Part I Control Unit Introduction Model 5000B Other NIR Absorbers that can be measured consult factory Introduction Model 5000B Main Features of the Analyzer 1 UP/DOWN Switch ESCAPE/ENTER SwitchOperator Interface Model 5000B Controls, Indicators, and Connectors Control Section Interface Panel Model 5000B Interface Panel Introduction Model 5000B Installation Unpacking the Control/Analysis UnitPart Primary Input Power Connections 115VACExamples Analog Output ConnectionsAlarm Relays Range Voltage Current mATypes of Relay Contacts The connectors are Part I RS-232 Signals RS-232 Sig RS-232 Pin PurposeTesting the System Command DescriptionParameter Setting Part Operation Photometric Analyzer Operation /Control UnitOperation/Control Unit Hierarchy of System Functions and Subfunctions Setting up an AUTO-CAL System FunctionSelect More and press the Enter Key PasswordProtection Entering the PasswordInstalling or Changing the Password Logging Out System Self-Diagnostic Test Model Screen CheckingLinearitywithALGORITHMDigital Filter Setup Setting 90% Response time secondsFilter or Solenoid Setup Hold/Track Setup Calibration/Hold Timer SetupCalbrt hold 3 min Sample hold 1 min Analog 4 to 20 mA Output CalibrationUse UP/DOWN arrow to Adjust 4 ma 250 ModelAuto Mode Zeroing Show NegativeZero Cal Manual Mode Zeroing Cell Failure SccmAuto Mode Spanning Manual Mode SpanningSpan Cal How to access the offset function Offset FunctionSelect zero mode Auto Select zero mode MAN Zero off 0.0 ppm ENT to begin ZeroAlarms Function AL1 1ØØØ ppm HI DftN FsN LtchN Range Select Function Manual Select/Define Range ScreenAuto Screen Precautions 2illustratestheseschemesgraphically Analyze Function Set Range Screen Algorithm Application More Curve Algorithm Screen Auto Mode Linearization Fuse Replacement Part I Control Unit/Analysis Unit MaintenancePower System Self Diagnostic TestMaintenance Model 5000B Photometric Analyzer AnalogPreamp Cell detectorPart Part II Part II Analysis UnitRoutine Maintenance Part II Analysis Unit Appendix Iv Part Operations/Analysis Unit Control Functions Explosion-ProofVersionControlModulePreliminaryInspection Control SettingsStart-p Analyzers not equipped with Auto ZeroPower On Observations Pre-Start-upElectricalCheckoutTeledyneAnalyticalInstruments4-5 Procedure CalibrationStandardizationFluids TeledyneAnalyticalInstruments4-7 Operations/Analysis Unit TeledyneAnalyticalInstruments4-9 Operations/Analysis Unit TeledyneAnalyticalInstruments4-11 Operations/Analysis Unit TeledyneAnalyticalInstruments4-13 Ppm simulated water using Span Zero SampleAttachment 5000B/5020 NIR Analyzer Operations/Analysis Unit Operational Theory Contd Source Module OpticalSystem Sample cellAnalyzerSystem-BlockDiagram Power Module Block Diagram Detectorcompartment System Visual Check and Response Procedure Routine MaintenancePart II Maintenance/Analysis Unit Automatic operation and routine operational dutiesSuggested Preventive Maintenance Schedule DailyEquipment Required Oscilloscope Display of the I to E Converter Output Part II Maintenance/Analysis Unit Setup of the Logarithmic Amplifier Inverting AmplifierBattery-Powered Oscilloscope Synchronization Point Interface Board Terminals StripPart Part II Maintenance/Analysis Unit Part Specifications Operating Temperature 0-50oCPhotometric Analyzer, 5000B Appendix Appendix Models 5000B Other Features Recommended 2-Year Spare Parts List QtyP/NDescriptionDrawing List, 5000B Appendix Models 5000B Teledyne Analytical Instruments