Teledyne M5000B operating instructions Auto Mode Linearization

Page 60

3 Operation/Control Unit

Model 5000B

 

 

 

 

 

 

 

 

 

Repeat the above procedure for each of the data points you are setting (up to nine points: 0-8). Set the points in unit increments. Do not skip numbers. The linearizerwillautomaticallyadjustforthenumberofpointsentered.

When you are done, Press ESCAPE. The message, Completed. Wait for calculation, appears briefly, and then the main System screen returns.

To end the session, send:

st<enter> st<enter>

to the analyzer from the computer.

3.8.2.2 Auto Mode Linearization

To linearize in the Auto Mode, you must have on hand a separate

calibration fluid for each of the data points you are going use in your linear- ization. First, the analyzer is zeroed and spanned as usual. Then, each special calibration fluid, for each of the intermediate calibration points, is flowed, in turn, through the sensor. As each fluid flows, the differential value for that intermediate calibration point is entered from the front panel of the analyzer.

Note: The span fluid used to span the analyzer must be >90% of the range being analyzed.

Before starting linearization, perform a standard calibration. See section 4.4. To enter data:

From the System Functions screen—

1.Use < > to select ALGORITHM , and Enter.

2.Select and Enter SETUP.

3.Enter AUTO from the Calibration Mode Select screen. The Auto Linearize Mode data entry screen appears.

19.5ppm SO2 Input(Ø) :20.00

5.Use the DÑ keys to set the proper value of calibration fluid, and Enter. Repeat this step for each cal-point number as it appears in the Input (x) parentheses.

6.Repeat step 5 for each of the special calibration fluides, from the lowest to the highest concentrations. Press Escape when done.

To end the session, send:

st<enter> st<enter>

to the analyzer from the computer.

3-32 Part I

Teledyne Analytical Instruments

Image 60
Contents Photometric Analyzer Teledyne Analytical InstrumentsModel 5000B Copyright 1999 Teledyne Analytical InstrumentsTable of Contents IiiModel 5000B Teledyne Analytical Instruments Part I Part I Control UnitInstallation Ii PartOperation Maintenance Iv Part Model 5000B Photometric AnalyzerOverview Typical ApplicationsPhotometric Analyzer Part I Control Unit Introduction Model 5000B Other NIR Absorbers that can be measured consult factory Introduction Model 5000B Main Features of the Analyzer 1 UP/DOWN Switch ESCAPE/ENTER SwitchOperator Interface Model 5000B Controls, Indicators, and Connectors Control Section Interface Panel Model 5000B Interface Panel Introduction Model 5000B Installation Unpacking the Control/Analysis UnitPart Primary Input Power Connections 115VACExamples Analog Output ConnectionsAlarm Relays Range Voltage Current mATypes of Relay Contacts The connectors are Part I RS-232 Signals RS-232 Sig RS-232 Pin PurposeTesting the System Command DescriptionParameter Setting Part Operation Photometric Analyzer Operation /Control UnitOperation/Control Unit Hierarchy of System Functions and Subfunctions Setting up an AUTO-CAL System FunctionSelect More and press the Enter Key PasswordProtection Entering the PasswordInstalling or Changing the Password Logging Out System Self-Diagnostic Test Model Screen CheckingLinearitywithALGORITHMDigital Filter Setup Setting 90% Response time secondsFilter or Solenoid Setup Hold/Track Setup Calibration/Hold Timer SetupCalbrt hold 3 min Sample hold 1 min Analog 4 to 20 mA Output CalibrationUse UP/DOWN arrow to Adjust 4 ma 250 ModelAuto Mode Zeroing Show NegativeZero Cal Manual Mode Zeroing Cell Failure SccmAuto Mode Spanning Manual Mode SpanningSpan Cal How to access the offset function Offset FunctionSelect zero mode Auto Select zero mode MAN Zero off 0.0 ppm ENT to begin ZeroAlarms Function AL1 1ØØØ ppm HI DftN FsN LtchN Range Select Function Manual Select/Define Range ScreenAuto Screen Precautions 2illustratestheseschemesgraphically Analyze Function Set Range Screen Algorithm Application More Curve Algorithm Screen Auto Mode Linearization Fuse Replacement Part I Control Unit/Analysis Unit MaintenanceSystem Self Diagnostic Test Maintenance Model 5000B Photometric AnalyzerPower AnalogPreamp Cell detectorPart Part II Part II Analysis UnitRoutine Maintenance Part II Analysis Unit Appendix Iv Part Operations/Analysis Unit Control Functions Explosion-ProofVersionControlModuleControl Settings Start-pPreliminaryInspection Analyzers not equipped with Auto ZeroPower On Observations Pre-Start-upElectricalCheckoutTeledyneAnalyticalInstruments4-5 Procedure CalibrationStandardizationFluids TeledyneAnalyticalInstruments4-7 Operations/Analysis Unit TeledyneAnalyticalInstruments4-9 Operations/Analysis Unit TeledyneAnalyticalInstruments4-11 Operations/Analysis Unit TeledyneAnalyticalInstruments4-13 Ppm simulated water using Span Zero SampleAttachment 5000B/5020 NIR Analyzer Operations/Analysis Unit Operational Theory Contd Source Module OpticalSystem Sample cellAnalyzerSystem-BlockDiagram Power Module Block Diagram Detectorcompartment Routine Maintenance Part II Maintenance/Analysis UnitSystem Visual Check and Response Procedure Automatic operation and routine operational dutiesSuggested Preventive Maintenance Schedule DailyEquipment Required Oscilloscope Display of the I to E Converter Output Part II Maintenance/Analysis Unit Setup of the Logarithmic Amplifier Inverting AmplifierBattery-Powered Oscilloscope Synchronization Point Interface Board Terminals StripPart Part II Maintenance/Analysis Unit Part Specifications Operating Temperature 0-50oCPhotometric Analyzer, 5000B Appendix Appendix Models 5000B Other Features Recommended 2-Year Spare Parts List QtyP/NDescriptionDrawing List, 5000B Appendix Models 5000B Teledyne Analytical Instruments