Emerson 54e pH/ORP Alarm setpoints Output setpoints, Simulated tests, Test output

Page 35

MODEL 54e pH/ORP

SECTION 5.0

 

SOFTWARE CONFIGURATION

5.4 TESTING OUTPUTS AND ALARMS

Alarm setpoints

Output setpoints

Simulated tests

ExitEnter

This section describes how the current outputs and alarm relays can be manually set for the purposes of checking devices such as valves, pumps, or recorders.

1.From the main menu, move the cursor down to "Program" and press Enter 4. On this display, move the cursor to "Simulated tests" and press Enter 4.

Test output 1

Test output 2

Test alarm 1

ExitEnter

Test alarm 2

Test alarm 3

Test alarm 4

ExitEnter

2.At this point there are six separate screens for testing each of the current outputs and each of the alarm relays. Highlight the desired item by pressing the arrow b t keys as needed.

When the desired item is highlighted, press Enter 4 to continue. Go to step 3a for outputs and 3b for alarms.

NOTE

A cautionary message will appear to warn that the out- put or alarm that was selected will be changed by the fol- lowing action. Be sure to alert plant personnel that these changes are simulated and do not represent a change in the actual process. Press Enter 4 to contin- ue or Abort 1 to cancel the simulation.

Test output 1: 10.00 mA

Simulating output1

ExitEdit

Test output 1: 10.00 mA

3a. The output is now being simulated. In the example to the left, out- put 1 has been set to 10.00 mA. The output will remain at 10.00 mA until either Exit 1 (or Edit 4 see below) is pressed or the test is concluded by timeout. The default value for the timeout is 10 min- utes, so after 10 minutes, the output would go back to normal oper- ation. To configure the timeout option, see Section 5.5.

If the displayed current is not the desired value, press the Edit 4 key and the next screen will allow changing the value. Use the arrow keys to change the display as needed, and press Test 4 to use that value. Press Esc 3 to cancel the change in the value and contin- ue simulating the previous current.

Esc Test

29

Image 35
Contents Model 54e pH/ORP PH/ORP Hart Analyzer/ControllerElectrical Shock Hazard Essential InstructionsPage Section Title Description and SpecificationsModel 54E PH/ORP Microprocessor Analyzer Table No Title Appendix TitleList of Tables About This Document Table of Contents List of FiguresDescription of Controls Section Description and SpecificationsGeneral Description Current Outputs Specifications Physical Specifications GeneralPower Relative Humidity 95%, non condensing AlarmsRecommended Sensors Analyzer Specifications @ 25CPart NO. Description Ordering InformationCode Options Unpacking and Inspection Section InstallationMechanical Installation Locating the ControllerPanel Mounting Alarm Relay Output Wiring Section WiringPower Input Wiring Analog Output Wiring454EPH02 Tions will Result in Controller MALFUNC- Tion PH Sensor WiringPH Sensor Compatibility 454EPH01 Cable Dressing Final Electrical Check Wiring to Model Introduction Section CalibrationTemperature Calibration Hold Mode OffAdjust temp 25.1 C Automatic TWO-POINT Calibration Autocal bufferBuffer 1 -- Wait Calibration Notes Auto buffer cal doneUse the arrow keys to select the correct buffer Manual TWO-POINT Calibration Calibrate pointPt 04.00pH 25C Pt calibration doneSINGLE-POINT pH Calibration Standardize 7.01 pH59.16 mV/pH Temperature Compensation OptionsTemp comp Auto Output trimTrim Outputs Hold ModeOutput trim PH Settings List Section Software ConfigurationRanges Factory Settings User Settings Diagnostics Section Outline of Menu Levels Setpoint 07.00 pH Alarm High 14.00pHChanging Alarm Setpoints Simulate testsChanging Output Setpoints PID only Setpoint 7.00 pH4mA 0.00 pH 20mA 2.00 pH Setpoint + 06.90 pHMA +00.00 pH OutputOutput MA 0.00 pH 20 mA 14.00 pH Output 1 12.00 mASimulated tests Alarm setpoints Output setpointsTest output Test alarm Testing Outputs and AlarmsTest alarm 1 Open Test alarm 1 OpenMeasure pH Display right AL2Security Caution Display contrast Timeout OnOutput Control Parameters Control ModeDisplay Changing Output ParametersRange 4-20 mA Fault 22.00 mARange 4-20 mAl Hold Disable feature Fault 22.00mAOutput 2 control Output 2 setup Hold feature setupInterval timer AlarmsAlarm 3 control Relay Default None Alarm LowSetpoint 0.01 pH Delay 0 sec Hysteresis 0.01 pH Delay 0 secURV 2.00 pH Relay default NoneSetpoint +07.00 pH Time period 30 secAlarm 4 Setup Alarm FaultFeed limit Disable Timeout 3600 secOn time 120 sec Timer DisableRecovery 600 sec Interval Timer Examples Zero offset 60 mV DiagnosticsDiagnostics Off Stabilize pH 0.01 pH Stabilize time 10 sec DiagnosticsAuto Calibration Setup Autocal StandardStandard Buffers Auto Calibration SetupSection Theory of Operation Continuous Sensor DiagnosticsSee .8 for instructions on chang- ing these setpoints Interval TimerTypical set points Time Proportional Control TPC Mode Code Alarm RelaysAnalog Outputs Normal ModeAction Definitions Controller Mode PriorityController Mode Priority Chart Condition DefinitionsMeasurement and Set Point Feedback Control PID Control Mode CombinationsPID Control Code PID ControlControl Loop Adjustment and Tuning Proportional Mode GainProportional Gain Plus Integral Reset Process Reaction Curve Process Reaction Curve MethodPID Control Level 1 3 Password Privileges Section Special Procedures and FeaturesPassword Protection Lock all Configuring SecuritySecurity Custom CurveSensor iso 07.00 pH Solution Temperature CompensationTemp coeff +.000 Operate iso +07.00 pH Troubleshooting is easy as 1, 2, 3… Section TroubleshootingDiagnostic Messages Description of problem Diagnostic MessagesSymptom Action Quick Troubleshooting GuideDiagnostic Variables Theoretical pH vs. Millivolt Values at 25C 77F Ohms Ambient Temperature Troubleshooting GuidelinesTemperature Compensation Circuit Troubleshooting ProcedureController preamp check Figure Preamplifier Troubleshooting ProcedurePreamp in a junction box Figure Junction Box Preamplifier Check No Power Troubleshooting GuideCoated glass electrode Clean with soft cloth and clean water Password protected Part Number Description Replacement PartsGeneral Warranty RepairNON-WARRANTY Repair Section Return of MaterialQuick Startup ORP Appendix a ORP ConfigurationItem Choices ORP Settings User Settings Table A-1. ORP Settings ListDiagnostics Section Choices ORP Settings User SettingsSimulated tests Output setpointsORP Calibration Standardize 400 mVSensor or Circuit Board only Return of Materials RequestReturn of Material WarrantySpecifications subject to change without notice