MODEL 54e pH/ORP |
| SECTION 5.0 | |
|
|
| SOFTWARE CONFIGURATION |
5.3 CHANGING OUTPUT SETPOINTS (NORMAL ONLY) | |||
Alarm setpoints |
| This section describes how the 4 (or 0) to 20 mA current outputs can be | |
| reranged. Note that the current outputs can be configured to represent | ||
|
|
| pH, temperature, glass impedance, or reference impedance. See |
| Output setpoints |
| |
|
| Section 5.6 for details on configuration. | |
Simulated test |
| 1. From the main menu, move the cursor down to "Program" and press | |
|
|
| Enter (F4). On this display, move the cursor to "Rerange outputs" |
| Exit | Enter | and press Enter (F4). |
Output 1
Output 2
2.Select the desired output by moving the cursor down to highlight it. When the correct output is highlighted, press Enter (F4) to get to the adjustment screen.
ExitEnter
CAUTION: Current Output 1 will be affected.
ExitEnter
3.This message asks for confirmation of the requested change. Changes in these settings may degrade process control, so use cau- tion when making changes. Press Enter (F4) to continue. Otherwise press Abort (F1).
4 mA: 0.00 pH
20 mA: 14.00 pH
Output 1: 12.00 mA
ExitEnter
4 mA: +00.00 pH
20 mA: 14.00 pH
Output 1: 12.00 mA
This screen allows changing the setpoints for output 1. A similar screen is available for output 2. The live current output now being transmitted by the controller is shown on the third line.
4.Press Edit (F4) to make changes in the setpoints. The Edit key changes to a Save key and the F3 key becomes active as an Esc key. Use the arrow keys to make the display read the desired val- ues for the high and low current output limits. When done, press Save (F4) to enter the changes into memory. Press Esc (F3) to can- cel changes.
NOTE
Outputs that have been configured as
Esc Save