Emerson 54e pH/ORP Auto buffer cal done, Use the arrow keys to select the correct buffer

Page 23

MODEL 54e pH/ORP

SECTION 4.0

 

CALIBRATION

4.02pH

Buf1 done: 4.01 @25°C

3.A screen like the one on the left will appear after stabilization.

Use the arrow keys to select the correct buffer.

If the correct buffer does not appear, refer to Section 5.9 and select a buffer group that includes the desired buffers. As usual, to cancel the calibration, press Abort (F1).

Abort Cont

When the correct buffer is shown, remove the sensor from buffer 1, rinse and gently dry it. Press Cont (F3) to continue.

NOTE

The calibration calculation does not occur until buffer 2 is com- pleted. Aborting the calibration at this point will not change the reading of the controller.

10.02pH

Autocal buffer 2

Abort Cont

4.This screen appears immediately after pressing the Cont key in step 3.

Place the sensor into the second buffer. Press Cont (F3) to proceed. The word "Wait" flashes (as before) until the sensor is stabilized.

Press Abort (F1) to cancel the calibration if the process is taking too long. (see step 2).

10.02pH

Buf2 done: 10.01 @25°C

Abort Cont

10.02pH

Auto buffer cal done

Abort Cont

5.A screen like the one on the left will appear after stabilization. Use the arrow keys to select the correct buffer, and press Cont (F3). If the correct buffer does not appear, refer to Section 5.9 and select a buffer group that includes the desired buffers. As usual, to cancel the calibration, press Abort (F1).

This completes the calibration. The controller has now recalculated the slope and the zero offset of the sensor.

CALIBRATION NOTES

1.A two-point calibration should always be followed by a sin- gle-point calibration (see Section 4.4).

2.After a buffer calibration, the electrode slope value can be checked to see if the electrode is aging.

3.If the sensor is not at the same temperature as the buffer the cal- ibration will be in error.

NOTE

If hold mode was turned ON, be certain to install the sensor back in the process and change the setting to OFF to resume normal operation before leaving the controller. The hold mode screen will appear again before the main display is shown. Follow the same routine as in the Note for step 1 to turn the Hold Mode Off and then press Exit (F1).

17

Image 23
Contents Model 54e pH/ORP PH/ORP Hart Analyzer/ControllerElectrical Shock Hazard Essential InstructionsPage Section Title Description and SpecificationsModel 54E PH/ORP Microprocessor Analyzer Table No Title Appendix TitleList of Tables About This Document Table of Contents List of FiguresDescription of Controls Section Description and SpecificationsGeneral Description Current Outputs Specifications Physical Specifications GeneralPower Relative Humidity 95%, non condensing AlarmsRecommended Sensors Analyzer Specifications @ 25CPart NO. Description Ordering InformationCode Options Unpacking and Inspection Section InstallationMechanical Installation Locating the ControllerPanel Mounting Alarm Relay Output Wiring Section WiringPower Input Wiring Analog Output Wiring454EPH02 Tions will Result in Controller MALFUNC- Tion PH Sensor WiringPH Sensor Compatibility 454EPH01 Cable Dressing Final Electrical Check Wiring to Model Introduction Section CalibrationTemperature Calibration Hold Mode OffAdjust temp 25.1 C Automatic TWO-POINT Calibration Autocal bufferBuffer 1 -- Wait Calibration Notes Auto buffer cal doneUse the arrow keys to select the correct buffer Manual TWO-POINT Calibration Calibrate pointPt 04.00pH 25C Pt calibration doneSINGLE-POINT pH Calibration Standardize 7.01 pH59.16 mV/pH Temperature Compensation OptionsTemp comp Auto Output trimTrim Outputs Hold ModeOutput trim PH Settings List Section Software ConfigurationRanges Factory Settings User Settings Diagnostics Section Outline of Menu Levels Setpoint 07.00 pH Alarm High 14.00pHChanging Alarm Setpoints Simulate testsChanging Output Setpoints PID only Setpoint 7.00 pH4mA 0.00 pH 20mA 2.00 pH Setpoint + 06.90 pHMA +00.00 pH OutputOutput MA 0.00 pH 20 mA 14.00 pH Output 1 12.00 mASimulated tests Alarm setpoints Output setpointsTest output Test alarm Testing Outputs and AlarmsTest alarm 1 Open Test alarm 1 OpenMeasure pH Display right AL2Security Caution Display contrast Timeout OnOutput Control Parameters Control ModeDisplay Changing Output ParametersRange 4-20 mA Fault 22.00 mARange 4-20 mAl Hold Disable feature Fault 22.00mAOutput 2 control Output 2 setup Hold feature setupInterval timer AlarmsAlarm 3 control Relay Default None Alarm LowSetpoint 0.01 pH Delay 0 sec Hysteresis 0.01 pH Delay 0 secURV 2.00 pH Relay default NoneSetpoint +07.00 pH Time period 30 secAlarm 4 Setup Alarm FaultFeed limit Disable Timeout 3600 secOn time 120 sec Timer DisableRecovery 600 sec Interval Timer Examples Zero offset 60 mV DiagnosticsDiagnostics Off Stabilize pH 0.01 pH Stabilize time 10 sec DiagnosticsAuto Calibration Setup Autocal StandardStandard Buffers Auto Calibration SetupSection Theory of Operation Continuous Sensor DiagnosticsSee .8 for instructions on chang- ing these setpoints Interval TimerTypical set points Time Proportional Control TPC Mode Code Alarm RelaysAnalog Outputs Normal ModeAction Definitions Controller Mode PriorityController Mode Priority Chart Condition DefinitionsMeasurement and Set Point Feedback Control PID Control Mode CombinationsPID Control Code PID ControlControl Loop Adjustment and Tuning Proportional Mode GainProportional Gain Plus Integral Reset Process Reaction Curve Process Reaction Curve MethodPID Control Level 1 3 Password Privileges Section Special Procedures and FeaturesPassword Protection Lock all Configuring SecuritySecurity Custom CurveSensor iso 07.00 pH Solution Temperature CompensationTemp coeff +.000 Operate iso +07.00 pH Troubleshooting is easy as 1, 2, 3… Section TroubleshootingDiagnostic Messages Description of problem Diagnostic MessagesSymptom Action Quick Troubleshooting GuideDiagnostic Variables Theoretical pH vs. Millivolt Values at 25C 77F Ohms Ambient Temperature Troubleshooting GuidelinesTemperature Compensation Circuit Troubleshooting ProcedureController preamp check Figure Preamplifier Troubleshooting ProcedurePreamp in a junction box Figure Junction Box Preamplifier Check No Power Troubleshooting GuideCoated glass electrode Clean with soft cloth and clean water Password protected Part Number Description Replacement PartsGeneral Warranty RepairNON-WARRANTY Repair Section Return of MaterialQuick Startup ORP Appendix a ORP ConfigurationItem Choices ORP Settings User Settings Table A-1. ORP Settings ListDiagnostics Section Choices ORP Settings User SettingsSimulated tests Output setpointsORP Calibration Standardize 400 mVSensor or Circuit Board only Return of Materials RequestReturn of Material WarrantySpecifications subject to change without notice