Emerson 54e pH/ORP instruction manual Range 4-20 mAl, Fault 22.00 mA

Page 40

MODEL 54e pH/ORP

SECTION 5.0

 

SOFTWARE CONFIGURATION

5.6 CHANGING OUTPUT PARAMETERS (continued)

Output Setup Parameters (for Normal Outputs)

Range: 4-20 mAl

Dampen: 0 sec

Hold: Last Value

ExitEdit

Hold: Last Value

Fixed Hold: 21.00 mA

Fault: 22.00 mA

ExitEdit

4a.

Menu Item

 

Options

 

 

 

 

 

Range

 

4-20mA/0-20mA

 

Dampen(ing)

 

0-299 sec

 

Hold

Last value/fixed value

 

Fixed Hold (if Output 1 hold is "fixed value")

0-22.00 mA

 

Fault (fixed value in a fault condition)

 

0-22.00 mA

 

 

 

 

These parameters can be adjusted by highlighting the desired item and pressing the Edit 4 key. Once Edit has been pressed, change the item as needed and then press Save 4 to store the value.

"Range" determines whether the 4-20mA or 0-20 mA convention is used for the current output. If the range is changed, be sure to rerange the outputs as described in Section 5.3.

"Dampening" is used to time-average the current output, smoothing out the effect of a noisy reading. Higher values provide more smooth- ing.

Enabling the "hold" feature will give the user the option of placing the output in hold during the calibration sequence.

A "fixed value" places the held output at a fixed value between 0 and 22 mA.

Output Setup Parameters (for PID Outputs only)

Setpoint: 7.00 pH

Proportional: 100.0%

Integral: 0 sec

ExitEdit

Derivative: 0.0%

Range: 4-20 mA

Hold: Last Value

ExitEdit

4b.

Menu Item

Options

 

Setpoint

-2 to 16 (pH), ±1400 mV (ORP/Redox)

 

 

-15 to 120°C (Temp), 0-140 kΩ(Ref Imp),

 

 

0-2000 MΩ(Glass Imp)

 

Proportional

0-299.9 %

 

Integral

0-2999 sec

 

 

 

 

Derivative

0-299.9 %

 

 

 

The four parameters above are only available for outputs that have been configured as PID outputs in step 3. These parameters can be adjusted using the same technique as in step 4a, by highlighting the desired item and pressing the Edit 4 key. Once Edit has been pressed, change the item as needed and then press Save 4 to store the value.

Use caution in changing the values of these parameters.

"Setpoint" is usually the desired value at which the process is being controlled, typically the output will be 4(or 0) mA when the parameter is near the setpoint. This setting can also be changed using the pro- cedure in Section 5.1.

34

Image 40
Contents PH/ORP Hart Analyzer/Controller Model 54e pH/ORPEssential Instructions Electrical Shock HazardPage Model 54E PH/ORP Microprocessor Analyzer Description and SpecificationsSection Title List of Tables Appendix TitleTable No Title Table of Contents List of Figures About This DocumentGeneral Description Section Description and SpecificationsDescription of Controls Specifications Physical Specifications General PowerRelative Humidity 95%, non condensing Alarms Current OutputsAnalyzer Specifications @ 25C Recommended SensorsCode Options Ordering InformationPart NO. Description Section Installation Mechanical InstallationLocating the Controller Unpacking and InspectionPanel Mounting Section Wiring Power Input WiringAnalog Output Wiring Alarm Relay Output Wiring454EPH02 PH Sensor Compatibility PH Sensor WiringTions will Result in Controller MALFUNC- Tion 454EPH01 Cable Dressing Final Electrical Check Wiring to Model Section Calibration IntroductionAdjust temp 25.1 C Hold Mode OffTemperature Calibration Buffer 1 -- Wait Autocal bufferAutomatic TWO-POINT Calibration Use the arrow keys to select the correct buffer Auto buffer cal doneCalibration Notes Calibrate point Pt 04.00pH 25CPt calibration done Manual TWO-POINT CalibrationStandardize 7.01 pH SINGLE-POINT pH CalibrationTemperature Compensation Options Temp comp AutoOutput trim 59.16 mV/pHOutput trim Hold ModeTrim Outputs Section Software Configuration PH Settings ListRanges Factory Settings User Settings Diagnostics Section Outline of Menu Levels Alarm High 14.00pH Changing Alarm SetpointsSimulate tests Setpoint 07.00 pHSetpoint 7.00 pH 4mA 0.00 pH 20mA 2.00 pHSetpoint + 06.90 pH Changing Output Setpoints PID onlyOutput OutputMA 0.00 pH 20 mA 14.00 pH Output 1 12.00 mA MA +00.00 pHAlarm setpoints Output setpoints Test output Test alarmTesting Outputs and Alarms Simulated testsTest alarm 1 Open Test alarm 1 OpenDisplay right AL2 Measure pHDisplay contrast Timeout On Security CautionControl Mode DisplayChanging Output Parameters Output Control ParametersRange 4-20 mAl Fault 22.00 mARange 4-20 mA Fault 22.00mA Output 2 control Output 2 setupHold feature setup Hold Disable featureAlarm 3 control AlarmsInterval timer Alarm Low Setpoint 0.01 pH Delay 0 secHysteresis 0.01 pH Delay 0 sec Relay Default NoneRelay default None Setpoint +07.00 pHTime period 30 sec URV 2.00 pHAlarm Fault Feed limit DisableTimeout 3600 sec Alarm 4 SetupRecovery 600 sec Timer DisableOn time 120 sec Interval Timer Examples Diagnostics Off DiagnosticsZero offset 60 mV Diagnostics Auto Calibration SetupAutocal Standard Stabilize pH 0.01 pH Stabilize time 10 secAuto Calibration Setup Standard BuffersContinuous Sensor Diagnostics Section Theory of OperationTypical set points Interval TimerSee .8 for instructions on chang- ing these setpoints Alarm Relays Time Proportional Control TPC Mode CodeNormal Mode Analog OutputsController Mode Priority Controller Mode Priority ChartCondition Definitions Action DefinitionsPID Control Mode Combinations PID Control CodePID Control Measurement and Set Point Feedback ControlProportional Gain Plus Integral Reset Proportional Mode GainControl Loop Adjustment and Tuning Process Reaction Curve Method Process Reaction CurvePID Control Password Protection Section Special Procedures and FeaturesLevel 1 3 Password Privileges Configuring Security SecurityCustom Curve Lock allTemp coeff +.000 Solution Temperature CompensationSensor iso 07.00 pH Operate iso +07.00 pH Section Troubleshooting Troubleshooting is easy as 1, 2, 3…Diagnostic Messages Diagnostic Messages Description of problemQuick Troubleshooting Guide Symptom ActionDiagnostic Variables Theoretical pH vs. Millivolt Values at 25C 77F Troubleshooting Guidelines Temperature Compensation CircuitTroubleshooting Procedure Ohms Ambient TemperaturePreamp in a junction box Figure Preamplifier Troubleshooting ProcedureController preamp check Figure Junction Box Preamplifier Check Troubleshooting Guide No PowerCoated glass electrode Clean with soft cloth and clean water Password protected Replacement Parts Part Number DescriptionWarranty Repair NON-WARRANTY RepairSection Return of Material GeneralAppendix a ORP Configuration Quick Startup ORPTable A-1. ORP Settings List Item Choices ORP Settings User SettingsChoices ORP Settings User Settings Diagnostics SectionOutput setpoints Simulated testsStandardize 400 mV ORP CalibrationReturn of Materials Request Sensor or Circuit Board onlyWarranty Return of MaterialSpecifications subject to change without notice