MODEL 54e pH/ORP | SECTION 5.0 |
| SOFTWARE CONFIGURATION |
TABLE 5-1. pH Settings List (continued)
ITEM |
| RANGES | FACTORY SETTINGS | USER SETTINGS | ||
5. | Alarm 3 Control |
|
|
|
|
|
| (a) Activation Method |
| Process/Temp |
| Process | _______ |
| (b) Alarm 3 Control Mode |
| Normal/TPC |
| Normal | _______ |
6a. Alarm 3 Setup (Normal) |
|
|
|
|
| |
| (a) Configuration |
| Low alarm/High alarm/Off |
| High | _______ |
| (b) Hysteresis (deadband) |
| 0 - 5.00 pH |
| 0.01 pH | _______ |
| (e) Delay Time |
|
| 0 sec | _______ | |
| (d) Relay Fault |
| Open/Closed/None |
| None | _______ |
6b. Alarm 3 Setup (TPC) |
|
|
|
|
| |
| (a) Setpoint |
| 7 pH | _______ | ||
| (b) Proportional |
|
| 100.0% | _______ | |
| (c) Integral |
|
| 0 sec | _______ | |
| (d) Derivative |
|
| 0.0% | _______ | |
| (e) Time Period |
|
| 30 sec | _______ | |
| (f) LRV (100% On) |
| 2 pH | _______ | ||
| (g) URV (100% Off) |
| 0 pH | _______ | ||
| (h) Relay Fault |
| None/Open/Closed |
| None | _______ |
7. | Alarm 4 Control |
|
|
|
|
|
| (a) Alarm |
| Fault/Off |
| Fault | _______ |
8. | Feed Limit Timer |
|
|
|
|
|
| (a) Feed Limit | Disable/alarm 1/alarm 2/alarm 3 | Disable | _______ | ||
| (b) Timeout Value |
|
| 3600 sec | _______ | |
C. Alarms (Section 5.7) |
|
|
|
|
| |
9. | Interval Timer |
|
|
|
|
|
| (a) Timer (selection) | Disable/alarm 1/alarm 2/alarm 3 | Disable | _______ | ||
| (b) Timer (activation method) | Time activated/Imped. activated | Time activated | _______ | ||
| (c) Interval |
|
| 24.0 hr | _______ | |
| (d) Repeats |
|
| 1 | _______ | |
| (e) On Time |
|
| 120 sec | _______ | |
| (f) Off Time |
|
| 1 sec | _______ | |
| (g) Recovery |
|
| 600 sec | _______ | |
D. Diagnostics (Section 5.8) |
|
|
|
|
| |
1. | Diagnostics (Glass and Reference) |
| On/Off |
| 000 (no security) | _______ |
2. | Glass Imp(edance) High Setpoint |
| 000 (no security) | _______ | ||
3. | Glass Imp(edance) Low Setpoint |
|
|
|
| |
4. | Ref(erence) Imp(edance) High |
|
| 000 (no security) | _______ | |
5. | Zero Offset |
|
| 000 (no security) | _______ | |
6. | Cal(ibration) Warn(ing) |
|
|
|
| |
7. | Imped(ance Temperature) Comp(ensation) | On/Off |
| 000 (no security) | _______ | |
E. Auto Calibration (Section 5.9) |
|
|
|
|
| |
1. | Autocal (Buffer List Selection) | Manual/Standard/DIN 19267 |
| 000 (no security) | _______ | |
2. | Stabilize pH (Auto Calibration) |
|
| 000 (no security) | _______ | |
3. | Stabilization Time (Auto Calibration) |
|
| 000 (no security) | _______ | |
F. Security (Section 7.1) |
|
|
|
|
| |
1. | Lock all |
|
| 000 (no security) | _______ | |
2. | Lock Program (Lock all except Calibrate) |
| 000 (no security) | _______ | ||
3. | Lock Config. (Lock all except Calibrate, |
|
|
|
|
|
| Output setpoints (PID), Simulated Tests |
|
|
|
|
|
| Alarm Setpoints, and Rerange Outputs) |
|
| 000 (no security) | _______ | |
D. Solution Temperature Compensation (Section 7.3) |
|
|
| |||
1. Temp(erature) Coeff(icient) |
|
| 000 (no security) | _______ | ||
2. | Operate Iso(potential pH) |
|
| 000 (no security) | _______ | |
3. | Sensor Iso(potential) |
|
| 000 (no security) | _______ | |
|
|
|
|
|
|
|
By changing the standard output configuration, you can set up the Model 54e pH/ORP to perform a wide variety of control and monitoring tasks. The configuration procedures allow you to program the controller to meet the specific control and monitoring requirements of your particular plant. This is done by recording the desired configuration parameters on the List of Settings Form and then actually configuring them by using the keys on the controller front panel.