Emerson 54e pH/ORP instruction manual Diagnostics Section

Page 30

MODEL 54e pH/ORP

SECTION 5.0

 

SOFTWARE CONFIGURATION

TABLE 5-1. pH Settings List (continued)

ITEM

 

RANGES

FACTORY SETTINGS

USER SETTINGS

5.

Alarm 3 Control

 

 

 

 

 

 

(a) Activation Method

 

Process/Temp

 

Process

_______

 

(b) Alarm 3 Control Mode

 

Normal/TPC

 

Normal

_______

6a. Alarm 3 Setup (Normal)

 

 

 

 

 

 

(a) Configuration

 

Low alarm/High alarm/Off

 

High

_______

 

(b) Hysteresis (deadband)

 

0 - 5.00 pH

 

0.01 pH

_______

 

(e) Delay Time

 

0-99 sec

 

0 sec

_______

 

(d) Relay Fault

 

Open/Closed/None

 

None

_______

6b. Alarm 3 Setup (TPC)

 

 

 

 

 

 

(a) Setpoint

-2 to 16 pH or -15 to 130°C

 

7 pH

_______

 

(b) Proportional

 

0-299.9%

 

100.0%

_______

 

(c) Integral

 

0-2999 sec

 

0 sec

_______

 

(d) Derivative

 

0-299.9%

 

0.0%

_______

 

(e) Time Period

 

10-2999 sec

 

30 sec

_______

 

(f) LRV (100% On)

-2 to 16 pH or -15 to 130°C

 

2 pH

_______

 

(g) URV (100% Off)

-2 to 16 pH or -15 to 130°C

 

0 pH

_______

 

(h) Relay Fault

 

None/Open/Closed

 

None

_______

7.

Alarm 4 Control

 

 

 

 

 

 

(a) Alarm

 

Fault/Off

 

Fault

_______

8.

Feed Limit Timer

 

 

 

 

 

 

(a) Feed Limit

Disable/alarm 1/alarm 2/alarm 3

Disable

_______

 

(b) Timeout Value

 

0-10,800 sec

 

3600 sec

_______

C. Alarms (Section 5.7)

 

 

 

 

 

9.

Interval Timer

 

 

 

 

 

 

(a) Timer (selection)

Disable/alarm 1/alarm 2/alarm 3

Disable

_______

 

(b) Timer (activation method)

Time activated/Imped. activated

Time activated

_______

 

(c) Interval

 

0-999.9 hr

 

24.0 hr

_______

 

(d) Repeats

 

1-60

 

1

_______

 

(e) On Time

 

0-2999 sec

 

120 sec

_______

 

(f) Off Time

 

0-2999 sec

 

1 sec

_______

 

(g) Recovery

 

0-999 sec

 

600 sec

_______

D. Diagnostics (Section 5.8)

 

 

 

 

 

1.

Diagnostics (Glass and Reference)

 

On/Off

 

000 (no security)

_______

2.

Glass Imp(edance) High Setpoint

0-2000 MOhms (0 disables)

 

000 (no security)

_______

3.

Glass Imp(edance) Low Setpoint

 

0-900 MOhms (0 disables)

 

 

 

4.

Ref(erence) Imp(edance) High

 

0-140 KOhms (0 disables)

 

000 (no security)

_______

5.

Zero Offset

 

0-999 mV (0 disables)

 

000 (no security)

_______

6.

Cal(ibration) Warn(ing)

 

0-500% (0 disables)

 

 

 

7.

Imped(ance Temperature) Comp(ensation)

On/Off

 

000 (no security)

_______

E. Auto Calibration (Section 5.9)

 

 

 

 

 

1.

Autocal (Buffer List Selection)

Manual/Standard/DIN 19267

 

000 (no security)

_______

2.

Stabilize pH (Auto Calibration)

 

.01-.50 pH

 

000 (no security)

_______

3.

Stabilization Time (Auto Calibration)

 

0-30 sec

 

000 (no security)

_______

F. Security (Section 7.1)

 

 

 

 

 

1.

Lock all

 

000-999

 

000 (no security)

_______

2.

Lock Program (Lock all except Calibrate)

000-999

 

000 (no security)

_______

3.

Lock Config. (Lock all except Calibrate,

 

 

 

 

 

 

Output setpoints (PID), Simulated Tests

 

 

 

 

 

 

Alarm Setpoints, and Rerange Outputs)

 

000-999

 

000 (no security)

_______

D. Solution Temperature Compensation (Section 7.3)

 

 

 

1. Temp(erature) Coeff(icient)

 

-0.044 to 0.028pH/°C

 

000 (no security)

_______

2.

Operate Iso(potential pH)

 

-1.35 to 20.12 pH

 

000 (no security)

_______

3.

Sensor Iso(potential)

 

0-14 pH

 

000 (no security)

_______

 

 

 

 

 

 

 

By changing the standard output configuration, you can set up the Model 54e pH/ORP to perform a wide variety of control and monitoring tasks. The configuration procedures allow you to program the controller to meet the specific control and monitoring requirements of your particular plant. This is done by recording the desired configuration parameters on the List of Settings Form and then actually configuring them by using the keys on the controller front panel.

24

Image 30
Contents PH/ORP Hart Analyzer/Controller Model 54e pH/ORPEssential Instructions Electrical Shock HazardPage Description and Specifications Model 54E PH/ORP Microprocessor AnalyzerSection Title Appendix Title List of TablesTable No Title Table of Contents List of Figures About This DocumentSection Description and Specifications General DescriptionDescription of Controls Relative Humidity 95%, non condensing Alarms Specifications Physical Specifications GeneralPower Current OutputsAnalyzer Specifications @ 25C Recommended SensorsOrdering Information Code OptionsPart NO. Description Locating the Controller Section InstallationMechanical Installation Unpacking and InspectionPanel Mounting Analog Output Wiring Section WiringPower Input Wiring Alarm Relay Output Wiring454EPH02 PH Sensor Wiring PH Sensor CompatibilityTions will Result in Controller MALFUNC- Tion 454EPH01 Cable Dressing Final Electrical Check Wiring to Model Section Calibration IntroductionHold Mode Off Adjust temp 25.1 CTemperature Calibration Autocal buffer Buffer 1 -- WaitAutomatic TWO-POINT Calibration Auto buffer cal done Use the arrow keys to select the correct bufferCalibration Notes Pt calibration done Calibrate pointPt 04.00pH 25C Manual TWO-POINT CalibrationStandardize 7.01 pH SINGLE-POINT pH CalibrationOutput trim Temperature Compensation OptionsTemp comp Auto 59.16 mV/pHHold Mode Output trimTrim Outputs Section Software Configuration PH Settings ListRanges Factory Settings User Settings Diagnostics Section Outline of Menu Levels Simulate tests Alarm High 14.00pHChanging Alarm Setpoints Setpoint 07.00 pHSetpoint + 06.90 pH Setpoint 7.00 pH4mA 0.00 pH 20mA 2.00 pH Changing Output Setpoints PID onlyMA 0.00 pH 20 mA 14.00 pH Output 1 12.00 mA OutputOutput MA +00.00 pHTesting Outputs and Alarms Alarm setpoints Output setpointsTest output Test alarm Simulated testsTest alarm 1 Open Test alarm 1 OpenDisplay right AL2 Measure pHDisplay contrast Timeout On Security CautionChanging Output Parameters Control ModeDisplay Output Control ParametersFault 22.00 mA Range 4-20 mAlRange 4-20 mA Hold feature setup Fault 22.00mAOutput 2 control Output 2 setup Hold Disable featureAlarms Alarm 3 controlInterval timer Hysteresis 0.01 pH Delay 0 sec Alarm LowSetpoint 0.01 pH Delay 0 sec Relay Default NoneTime period 30 sec Relay default NoneSetpoint +07.00 pH URV 2.00 pHTimeout 3600 sec Alarm FaultFeed limit Disable Alarm 4 SetupTimer Disable Recovery 600 secOn time 120 sec Interval Timer Examples Diagnostics Diagnostics OffZero offset 60 mV Autocal Standard DiagnosticsAuto Calibration Setup Stabilize pH 0.01 pH Stabilize time 10 secAuto Calibration Setup Standard BuffersContinuous Sensor Diagnostics Section Theory of OperationInterval Timer Typical set pointsSee .8 for instructions on chang- ing these setpoints Alarm Relays Time Proportional Control TPC Mode CodeNormal Mode Analog OutputsCondition Definitions Controller Mode PriorityController Mode Priority Chart Action DefinitionsPID Control PID Control Mode CombinationsPID Control Code Measurement and Set Point Feedback ControlProportional Mode Gain Proportional Gain Plus Integral ResetControl Loop Adjustment and Tuning Process Reaction Curve Method Process Reaction CurvePID Control Section Special Procedures and Features Password ProtectionLevel 1 3 Password Privileges Custom Curve Configuring SecuritySecurity Lock allSolution Temperature Compensation Temp coeff +.000Sensor iso 07.00 pH Operate iso +07.00 pH Section Troubleshooting Troubleshooting is easy as 1, 2, 3…Diagnostic Messages Diagnostic Messages Description of problemQuick Troubleshooting Guide Symptom ActionDiagnostic Variables Theoretical pH vs. Millivolt Values at 25C 77F Troubleshooting Procedure Troubleshooting GuidelinesTemperature Compensation Circuit Ohms Ambient TemperaturePreamplifier Troubleshooting Procedure Preamp in a junction box FigureController preamp check Figure Junction Box Preamplifier Check Troubleshooting Guide No PowerCoated glass electrode Clean with soft cloth and clean water Password protected Replacement Parts Part Number DescriptionSection Return of Material Warranty RepairNON-WARRANTY Repair GeneralAppendix a ORP Configuration Quick Startup ORPTable A-1. ORP Settings List Item Choices ORP Settings User SettingsChoices ORP Settings User Settings Diagnostics SectionOutput setpoints Simulated testsStandardize 400 mV ORP CalibrationReturn of Materials Request Sensor or Circuit Board onlyWarranty Return of MaterialSpecifications subject to change without notice