MODEL 54e pH/ORP | SECTION 5.0 |
| SOFTWARE CONFIGURATION |
5.8 ON-LINE DIAGNOSTICS SETUP (continued)
"Cal warn" compares the
"Imped comp" is used to compensate the glass impedance for tempera- ture changes. This is especially important at temperature extremes.
NOTE
The actual
5.9 AUTO CALIBRATION SETUP
|
|
|
| This section describes how the controller may be set up for auto calibra- |
| Diagnostics |
| tion, which includes automatic buffer recognition and auto stabilization. | |
|
| Alternatively, the controller may be configured to perform a manual cali- | ||
| Auto Calibration |
|
| |
|
|
| bration without the automatic features. | |
| Security |
| 1. Beginning from the main menu, move the cursor down to "Program" | |
|
| and press Enter (F4). From the program menu, move the cursor | ||
|
|
|
| |
| Exit | Enter | down using the arrow key to highlight "Configure" and press Enter | |
| (F4). | |||
|
|
|
| |
|
|
|
| Use the arrow key again to highlight "Auto Calibration" (as shown on |
|
|
|
| the left) and press Enter (F4). |
|
|
|
| NOTE |
|
|
|
| There are three buffer groups that may be selected. Automatic |
|
|
|
| buffer recognition includes temperature curves from 0 to 60°C. |
|
|
|
| a. Standard: Includes NIST (US), BSM (Britain), JIS 8802 (Japan) and |
|
|
|
| DIN 19266 (Germany) buffers. See Table |
|
|
|
| b. DIN 19267: 1.09/4.65/6.79/9.23/12.75 |
|
|
|
| c. Manual: (for manual calibration) disables automatic buffer recognition |
|
|
|
| and stabilization. |
|
|
|
|
|
Autocal: Standard
Stabilize pH: 0.01 pH
Stabilize time: 10 sec
ExitEdit
2. | Menu Item | Options |
|
|
|
| Autocal | Standard/ DIN 19267/ Manual |
|
|
|
| Stabilize pH | |
|
|
|
| Stabilize time | |
|
|
|
The values now being used by the controller are displayed. To change any of these items, use the arrow key to highlight the desired item and press Edit(F4). Use the arrow keys to make the change and press Save (F4) to enter the change into memory.
The pH reading must not change by more than the stabilize pH value over the stabilize time to be considered stable. Using a smaller pH value and a larger time provides the best protection against calibration while the reading is still changing.
43