Emerson 54e pH/ORP instruction manual Analyzer Specifications @ 25C, Recommended Sensors

Page 9

MODEL 54e pH/ORP

SECTION 1.0

 

DESCRIPTION AND SPECIFICATIONS

The Model 54e when configured as a pH analyzer, requires a dual (glass and reference) impedance pre- amplifier (patent pending). This preamp converts the high impedance pH glass electrode signal to a low impedance signal. The preamplifier may be located in one of three areas: 1) in the pH sensor for best per- formance, 2) in a remote junction box when process temperatures exceed 80°C (176°F) in submersion applications, or 3) in the analyzer when the distance between the pH sensor and the analyzer is 4.5 meters (15 feet) or less.

The Model 54e pH measures over the full range of 0- 14 pH. The current output may be calibrated to repre- sent any 1 to 14 pH span.

A two-point calibration is made by immersing the sen- sor in two different buffer solutions and entering the pH values. When two buffers are used, the micro- processor automatically calculates the electrode slope which is used for self-diagnostics. The electrode slope can be read on the display and manually adjusted. A one-point process standardization is easily accom- plished by entering the pH value of a grab sample.

ANALYZER SPECIFICATIONS @ 25°C

Measurement Range: 0 to 14 pH

Output Scale Expansion: Zero suppression: up to 13 pH units

Span: Any pH from 1 to 14

Accuracy: ± 0.01 pH

Repeatability: ± 0.01 pH

Stability: ± 0.01 pH/month, non-cumulative

Temperature Coefficient: Input: ± 0.003 pH/°C

Output: ± 0.006 pH/°C

Temperature Compensation : Pt 100 or Pt 1000 RTD, Automatic or Manual

–15 to 120°C (5 to 248°F)

The Model 54e when configured as an ORP analyzer, measures over a range of -1400 mV to +1400 mV in either the American convention (Oxidation Reduction Potential), or the European con- vention (Reduction Oxidation-Redox). Although tem- perature compensation is not used for ORP measurements, the process temperature is measured and displayed. Temperature measurement is made by an RTD located in the sensor assembly.

ANALYZER SPECIFICATIONS @ 25°C

Measurement Range: –1400 to +1400 mV

Output Scale Expansion: Zero suppression: up to ±1300 mV

Span: Any ORP range from 100 to 2800 mV

Accuracy: ± 1.0 mV

Repeatability: ± 1.0 mV

Stability: ± 1.0 mV/month, non-cumulative

Temperature Coefficient: Input: ± 0.2 mV/°C

Output: ± 0.4 mV/°C

Temperature Measurement: –15 to 120°C (5 to 248°F) Pt 100 or Pt 1000 RTD

RECOMMENDED SENSORS:

Model 330 Flow Through ORP

*Model 371 EuroSenz ORP

Model 381+ Insertion/Submersion/Flow Through ORP *Model 389 Disposable ORP

Model 396P Disposable ORP

Model 396R Retractable ORP

Model 398 Insertion/Submersion ORP

Model 398VP Insertion/Submersion with VP 6.0 connector Model 398R Retractable ORP

Model 398RVP Retractable ORP with VP 6.0 connector

RECOMMENDED SENSORS:

Model 320B Flow Through pH

Model 320HP High Purity pH *Model 328A Steam Sterilizable pH *Model 370 and 371 EuroSenz pH

Model 381+ Insertion/Submersion/Flow Through pH

*Model 389 Disposable pH *Model 396/VP Disposable pH Model 396P/VP Disposable pH Model 396R/VP Retractable pH Model 397 Quik Disconnect pH

Model 398/VP Insertion/Submersion pH

*Model 398R/VP Retractable pH *Model 399 Disposable pH

*Model Hx338 Steam Sterilizable pH

*Model Hx348 Steam Sterilizable pH

The Model 54e when ordered for ISE capability,

is suitable for use with a number of ion-selective elec- trodes. Consult the factory for available measure- ments and ranges.

3

Image 9
Contents Model 54e pH/ORP PH/ORP Hart Analyzer/ControllerElectrical Shock Hazard Essential InstructionsPage Description and Specifications Model 54E PH/ORP Microprocessor AnalyzerSection Title Appendix Title List of TablesTable No Title About This Document Table of Contents List of FiguresSection Description and Specifications General DescriptionDescription of Controls Power Specifications Physical Specifications GeneralRelative Humidity 95%, non condensing Alarms Current OutputsRecommended Sensors Analyzer Specifications @ 25COrdering Information Code OptionsPart NO. Description Mechanical Installation Section InstallationLocating the Controller Unpacking and InspectionPanel Mounting Power Input Wiring Section WiringAnalog Output Wiring Alarm Relay Output Wiring454EPH02 PH Sensor Wiring PH Sensor CompatibilityTions will Result in Controller MALFUNC- Tion 454EPH01 Cable Dressing Final Electrical Check Wiring to Model Introduction Section CalibrationHold Mode Off Adjust temp 25.1 CTemperature Calibration Autocal buffer Buffer 1 -- WaitAutomatic TWO-POINT Calibration Auto buffer cal done Use the arrow keys to select the correct bufferCalibration Notes Pt 04.00pH 25C Calibrate pointPt calibration done Manual TWO-POINT CalibrationSINGLE-POINT pH Calibration Standardize 7.01 pHTemp comp Auto Temperature Compensation OptionsOutput trim 59.16 mV/pHHold Mode Output trimTrim Outputs PH Settings List Section Software ConfigurationRanges Factory Settings User Settings Diagnostics Section Outline of Menu Levels Changing Alarm Setpoints Alarm High 14.00pHSimulate tests Setpoint 07.00 pH4mA 0.00 pH 20mA 2.00 pH Setpoint 7.00 pHSetpoint + 06.90 pH Changing Output Setpoints PID onlyOutput OutputMA 0.00 pH 20 mA 14.00 pH Output 1 12.00 mA MA +00.00 pHTest output Test alarm Alarm setpoints Output setpointsTesting Outputs and Alarms Simulated testsTest alarm 1 Open Test alarm 1 OpenMeasure pH Display right AL2Security Caution Display contrast Timeout OnDisplay Control ModeChanging Output Parameters Output Control ParametersFault 22.00 mA Range 4-20 mAlRange 4-20 mA Output 2 control Output 2 setup Fault 22.00mAHold feature setup Hold Disable featureAlarms Alarm 3 controlInterval timer Setpoint 0.01 pH Delay 0 sec Alarm LowHysteresis 0.01 pH Delay 0 sec Relay Default NoneSetpoint +07.00 pH Relay default NoneTime period 30 sec URV 2.00 pHFeed limit Disable Alarm FaultTimeout 3600 sec Alarm 4 SetupTimer Disable Recovery 600 secOn time 120 sec Interval Timer Examples Diagnostics Diagnostics OffZero offset 60 mV Auto Calibration Setup DiagnosticsAutocal Standard Stabilize pH 0.01 pH Stabilize time 10 secStandard Buffers Auto Calibration SetupSection Theory of Operation Continuous Sensor DiagnosticsInterval Timer Typical set pointsSee .8 for instructions on chang- ing these setpoints Time Proportional Control TPC Mode Code Alarm RelaysAnalog Outputs Normal ModeController Mode Priority Chart Controller Mode PriorityCondition Definitions Action DefinitionsPID Control Code PID Control Mode CombinationsPID Control Measurement and Set Point Feedback ControlProportional Mode Gain Proportional Gain Plus Integral ResetControl Loop Adjustment and Tuning Process Reaction Curve Process Reaction Curve MethodPID Control Section Special Procedures and Features Password ProtectionLevel 1 3 Password Privileges Security Configuring SecurityCustom Curve Lock allSolution Temperature Compensation Temp coeff +.000Sensor iso 07.00 pH Operate iso +07.00 pH Troubleshooting is easy as 1, 2, 3… Section TroubleshootingDiagnostic Messages Description of problem Diagnostic MessagesSymptom Action Quick Troubleshooting GuideDiagnostic Variables Theoretical pH vs. Millivolt Values at 25C 77F Temperature Compensation Circuit Troubleshooting GuidelinesTroubleshooting Procedure Ohms Ambient TemperaturePreamplifier Troubleshooting Procedure Preamp in a junction box FigureController preamp check Figure Junction Box Preamplifier Check No Power Troubleshooting GuideCoated glass electrode Clean with soft cloth and clean water Password protected Part Number Description Replacement PartsNON-WARRANTY Repair Warranty RepairSection Return of Material GeneralQuick Startup ORP Appendix a ORP ConfigurationItem Choices ORP Settings User Settings Table A-1. ORP Settings ListDiagnostics Section Choices ORP Settings User SettingsSimulated tests Output setpointsORP Calibration Standardize 400 mVSensor or Circuit Board only Return of Materials RequestReturn of Material WarrantySpecifications subject to change without notice