MODEL 54e pH/ORP |
| SECTION 8.0 |
|
| TROUBLESHOOTING |
TABLE |
| |
|
|
|
PROBLEM OR CONDITION | PROBABLE CAUSE AND | FOR MORE HELP, |
| CORRECTIVE ACTION | REFER TO |
Display reads between 3 and | Electrode cracked - |
|
6 pH regardless of actual pH | Replace electrode. |
|
solution or sample. |
|
|
"cracked glass failure"
Sluggish or slow display indica- tion for real changes in pH level.
Inability to change parameters in the controller. "Level 1 security: Lock" "Level 2 security: Lock" "Level 3 security: Lock"
Outputs do not change "Hold Mode activated" "simulating output 1 or 2" "simulating alarm 1,2,3, or 4"
Red LED light is on Various fault messages shown on lower display.
Electrode coated -
Clean electrode as instructed in sensor manual
Password protected - | Section 7.0 |
Your controller has password protection. |
|
You must enter the correct password to |
|
make changes in the controller. |
|
Controller is in Hold or simulate mode - | Software Configuration, Section 5.6 |
To remove from Hold, press any key | Calibration , Section 4.7 |
and then press Enter (F4). Press Edit (F4), |
|
change "On" to Off", and press Save (F4). |
|
To stop simulation, press Exit (F1). |
|
When the unit has been put into "Hold", |
|
all outputs go to their default states. |
|
Unit has gone into fault mode - | Section 8.0 |
Read the message and take action as |
|
indicated in Table |
|
68