MODEL 54e pH/ORP |
| SECTION 5.0 | |||
|
|
|
|
| SOFTWARE CONFIGURATION |
5.6 CHANGING OUTPUT PARAMETERS | |||||
| Display |
| This section describes the options available for configuration of the out- | ||
|
| puts. Each output can be configured to represent pH (or ORP/Redox, see | |||
|
|
|
|
| Appendix A), temperature, glass impedance, or reference impedance. In |
| Outputs |
|
|
| |
|
|
|
| addition, each output can be linear (normal) or nonlinear (PID). Several | |
| Alarms |
| options for both modes of outputs can be configured. This section | ||
|
| describes the features and shows how to make modifications in the set- | |||
|
|
|
|
| |
| Exit | Enter | tings. | ||
|
| ||||
|
|
|
|
| 1. Beginning from the main menu, move the cursor down to "Program" |
|
|
|
|
| and press Enter 4. From the program menu, move the cursor down |
|
|
|
|
| using the arrow key b to highlight "Configure" and press Enter 4. |
|
|
|
|
| Use the arrow key again to highlight "Outputs" (as shown on the left) |
|
|
|
|
| and press Enter 4. |
|
|
|
|
|
|
| Output 1 control |
| 2. There are 5 menu headers that relate to outputs. Each output has a | ||
| Output 1 setup |
| control header and a setup header. In addition, the hold feature is | ||
|
| turned on here. The hold feature is mostly used to prevent upsets dur- | |||
| Output 2 control |
| ing buffer calibration of the pH sensor. | ||
|
|
| |||
|
|
|
|
| To access each header, highlight the desired item and press the Enter |
| Exit | Enter | 4 key. To select another header, use the arrow keys. The bottom | ||
|
|
|
|
| menu header will only be highlighted if the end of the menu has been |
|
|
|
|
| reached. |
Output 2 control Output 2 setup Hold feature setup
ExitEdit
NOTE
Always configure the control parameters BEFORE mak- ing changes in the output setup. Changes in the output setup in step 4 will depend on the options that have been selected in step 3.
Output Control Parameters |
|
|
| |||
3. | Menu Item | Options | ||||
| ||||||
| ||||||
|
|
|
|
|
| |
|
|
|
| Output Measurement | Process/Temperature/Ref | |
| Output Measurement |
|
| |||
|
|
|
| Impedance/Glass Impedance | ||
| Control Mode |
| ||||
|
|
| ||||
| Control Mode | Normal/PID | ||||
|
|
|
|
|
|
Exit | Enter | Each output can be configured with the options above. The default | |
options are that output 1 is Process (pH or ORP/Redox), output 2 is | |||
|
| ||
|
| Temperature and that both outputs are Normal (not PID). This is the | |
|
| most common configuration and may not require changes. If no | |
|
| changes are desired, skip to step 4. | |
Output: Process |
| ||
| To make changes in these parameters, highlight the desired menu | ||
|
| ||
|
| header and press Enter 4. The value now being used is displayed | |
|
| and the F4 key can be pressed to Edit the item. Once Edit has been | |
|
| pressed, change the item as needed and then press Save 4 to | |
Exit | Edit | store the value. Repeat for the other output and/or items as needed. | |
|
33