MODEL 54e pH/ORP | SECTION 5.0 |
| SOFTWARE CONFIGURATION |
SECTION 5.0
SOFTWARE CONFIGURATION
This section contains the following:
•An introduction to using the configuration process
•A List of Settings for the controller
•
INTRODUCTION TO CONFIGURATION
The controller arrives from the factory configured and ready to operate as a pH controller. Refer to Appendix A for ORP meas- urements.
Figure 5-1 is an outline of the menu structure. Before attempting any changes refer to the parameter setup list shown in Table 5-1. This table presents a brief description and the possible options.
The factory setting is listed with a space for the user setting. It is recommended that the list be carefully reviewed before any changes are made.
On initial configuration, it is recommended that the parameters be entered in the order shown on the worksheet. This will reduce the chance of accidentally omitting a needed parameter.
TABLE 5-1. pH Settings List
ITEM | CHOICES | FACTORY SETTINGS | USER SETTINGS | ||
PROGRAM LEVEL (Sections 5.1 - 5.3) |
|
|
|
| |
A. Alarm Setpoints (Section 5.2) |
|
|
|
| |
1. Alarm 1 (low action) | 0 - 14 pH |
| 0.00 pH | _______ | |
2. Alarm 2 (high action) | 0 - 14 pH |
| 14.00 pH | _______ | |
3. Alarm 3 (high action) | 0 - 14 pH |
| 14.00 pH | _______ | |
B. Output Setpoints (Section 5.1, 5.3) |
|
|
|
| |
1. | Output 1: 4 mA | 0 - 14 pH |
| 0.00 pH | _______ |
2. | Output 1: 20 mA | 0 - 14 pH |
| 14.00 pH | _______ |
3. | Output 2: 4 mA |
| 0.0°C | _______ | |
4. | Output 2: 20 mA |
| 100.0°C | _______ | |
CONFIGURE LEVEL (Sections |
|
|
|
| |
A. Display (Section 5.5) |
|
|
|
| |
1. | Measurement type | pH/ORP/Redox |
| pH | _______ |
2. pH Resolution | 0.01 pH/0.1 pH |
| 0.01 pH |
| |
3. Temperature Units | °C/°F |
| °C | _______ | |
4. | Output 1 Units | mA/% (of full scale) |
| mA | _______ |
5. | Output 2 Units | mA/% (of full scale) |
| mA | _______ |
6. | Language | English/Français/Español/Deutsch/Italiano | English | _______ | |
7. | Main display lower left | See Section 5.5 |
| Alarm 1 Setpoint | _______ |
8. | Main display lower right | See Section 5.5 |
| Alarm 2 Setpoint | _______ |
9. | Display contrast |
| 5 | _______ | |
10. Test Timeout | On/Off |
| On | _______ | |
11. Timeout Value |
| 10 min | _______ | ||
B. Outputs (Section 5.6) |
|
|
|
| |
1. | Output 1 Control |
|
|
|
|
| (a) Output 1 Measurement | Process/Temp/Glass Imp/Ref Imp | Process (pH) | _______ | |
| (b) Output1 Control Mode | Normal/PID |
| Normal | _______ |
2a. Output 1 Setup (Normal) |
|
|
|
| |
| (a) Current Range |
| _______ | ||
| (b) Dampening |
| 0 Sec | _______ | |
| (c) Hold Mode | Last value/Fixed value |
| Last value | _______ |
| (d) Fixed Hold Value (if (c) Fixed) |
| 21 mA | _______ | |
| (e) Fault value |
| 22 mA | _______ |
22Continued on the following page