7.0 Wiring

7.1.POWER

Power is supplied through a fused, AC jack on the rear panel (item 2).Use the power cord supplied with the unit (PN15-17-011 for 115 VAC, 60Hz). See the following table for selecting the proper fuse rating. Use a metric, 5 x 20 mm sized, time-delayed fuse.

Power Setting (50 – 60 Hz)

Fuse Rating

THI P/N

100 VAC

0.315 Amp/250 VAC

23-05-038

115 VAC

0.60 Amp/250 VAC

23-05-039

230 VAC

0.70 Amp/250 VAC

23-05-040

Cords without plugs are supplied with units shipped outside of the U.S. Consult and comply with any local laws and/or codes when connecting to any AC main. The AC input is user selectable between 100, 115 or 230 VAC, 50 or 60 Hz, via an AC selector switch next to the AC jack (Item 3).

WARNING: Be sure to set the power select switch prior to connecting to mains. Re-fuse the connector according to the table above.

7.2. COMMUNICATIONS

Connectors J6 (Item 4), J7 (Item 6) and J9 (Item 6) are for RS-232, RS-485 connections respectively. Settings for serial communication are accessible via the front panel.

J6

1

Unused

 

2

Tx

 

3

Rx

RS-232

4

Unused

5

Gnd

(DB-9)

6

Unused

 

7

RTS

 

8

CTS

 

9

Unused

J7, J9

1

Unused

 

2

Rx-

 

3

Tx+

RS-485

4

Unused/Gnd (Gnd)

5

Unused

(DB-9)

6

Gnd/VCC (VCC)

 

7

R+

 

8

T-

 

9

Unused

164-122012 PowerPOD 400

Page 12 of 42

Image 12
Contents Teledyne Hastings Manual Print History Table of Contents Four to Twenty Milliamp Operation Power Quick Start InstructionsSafety Features Front Panel LockoutPower Selection Specifications PowerPod-400 Specifications TableFront Panel Channel Number Select KeysRear Panel Wiring PowerCommunications Alarms Transducer ConnectionsAnalog Signal Followers Power ON/OFF Changing the Command SET PointManual Operation Override OpenSetting a Channel to Auto Control Setting a Channel to Directly Meter Incoming SignalsSetting a Channel to Display the Totalizer Function Mode Reset Value External/Remote Operation Choosing RS-232 CommunicationChoosing RS-485 Communication Setup/Cal Mode Entering the SETUP/CAL ModeDisplay Setup Blanking a Display Line/ChannelSetting A/D Conversion Rate Filter Selecting Unit-of-MeasureSelecting Gas ID Totalizer Setup Front Panel Lock OutCount Up to a Set Point External Communication Setup RS-232 SettingsRS-485 Settings Count Down from a Set PointSelecting the Analog Signal Level Zero to Five Volt OperationZero to Ten Volt Operation Four to Twenty Milliamp OperationSetting Limit Alarms Setting a Single Channel’s High LimitSetting a Single Channel’s Low Limit Setting a Single Channel’s HysteresisSetting Ratio Control Parameters Enabling Ratio ControlCalibrating a Channel to ITS Incoming Signals Resetting the Zero OnlyResetting the Span Only Resetting Zero and Span Setting a Multiplication FactorCommands Alarm/Flag QueriesSerial Communication Set Point QueriesPowerPOD PowerPod-400 Serial Commands Appendix aPowerPod-400 Serial Commands Contd Appendix a Notes ApppendixCB Name Rate TotalGas ID Table Page 1 Gas ID Table Page 2 Setting the Zero & Span PowerPod-400 Power Supply/Totalizer Appendix DZeroing Unit 4 to 20 mA Range Spanning Unit Transducer POWERPOD-400 PowerPOD Drawings Warranty Warranty Repair Policy