10.0 Setup/Cal Mode

10.1. ENTERING THE SETUP/CAL MODE

Pressing the MODE key causes the MODE menu to be displayed. Choose the SETUP/CAL mode by pressing the number “3” key then ENTER. A CHANNEL SELECT menu is presented on the display. It should be noted that, although the user must select a specific channel number, some choices in the following menus may affect the operation of all channels and/or the display (ex; Display Brightness, type of communication). See the following menu options for more details.

10.2. DISPLAY SETUP

After entering the SETUP/CAL mode and selecting a channel number as described in the previous section, several selections are available that will affect the display. Follow the steps below to gain the desired customized display.

Note: Front panel lockout and display line (channel) blanking are only available through serial communication.

10.2.1. Blanking a Display Line/Channel

Use the serial command “DndCr”, where n = channel number,

d = 1 for Totalizer Mode, 2 for Meter Mode,

3 for a blanked line and Cr = carriage return.

10.2.2. Selecting Display Brightness

While the SETUP/CAL menu is displayed, choose DISPLAY by pressing the keys “1” + ENTER. Choose BRIGHTNESS by pressing the number “1” key + ENTER. The BRIGHTNESS menu will be visible. One of four different brightness levels can be chosen by pressing a number key. The number “1” key selects the dimmest setting while the number “4” key selects the brightest. Any change in brightness will affect the entire display after pressing the ENTER key.

164-122012 PowerPOD 400

Page 19 of 42

Image 19
Contents Teledyne Hastings Manual Print History Table of Contents Four to Twenty Milliamp Operation Quick Start Instructions PowerSafety Front Panel Lockout FeaturesPower Selection PowerPod-400 Specifications Table SpecificationsChannel Number Select Keys Front PanelRear Panel Power WiringCommunications Transducer Connections AlarmsAnalog Signal Followers Override Open Power ON/OFFChanging the Command SET Point Manual OperationSetting a Channel to Directly Meter Incoming Signals Setting a Channel to Auto ControlSetting a Channel to Display the Totalizer Function Mode Reset Value Choosing RS-232 Communication External/Remote OperationChoosing RS-485 Communication Blanking a Display Line/Channel Setup/Cal ModeEntering the SETUP/CAL Mode Display SetupSelecting Unit-of-Measure Setting A/D Conversion Rate FilterSelecting Gas ID Front Panel Lock Out Totalizer SetupCount Up to a Set Point Count Down from a Set Point External Communication SetupRS-232 Settings RS-485 SettingsFour to Twenty Milliamp Operation Selecting the Analog Signal LevelZero to Five Volt Operation Zero to Ten Volt OperationSetting a Single Channel’s Hysteresis Setting Limit AlarmsSetting a Single Channel’s High Limit Setting a Single Channel’s Low LimitEnabling Ratio Control Setting Ratio Control ParametersResetting the Zero Only Calibrating a Channel to ITS Incoming SignalsResetting the Span Only Setting a Multiplication Factor Resetting Zero and SpanSet Point Queries CommandsAlarm/Flag Queries Serial CommunicationPowerPOD Appendix a PowerPod-400 Serial CommandsPowerPod-400 Serial Commands Contd Appendix a Notes Name Rate Total ApppendixCBGas ID Table Page 1 Gas ID Table Page 2 Appendix D Setting the Zero & Span PowerPod-400 Power Supply/TotalizerZeroing Unit 4 to 20 mA Range Spanning Unit Transducer POWERPOD-400 PowerPOD Drawings Warranty Repair Policy Warranty