Teledyne powerpod 400 Selecting the Analog Signal Level, Zero to Five Volt Operation

Page 23

The default address as programmed at the factory is 01. If a different address is required, it may be changed using either RS-232 or 485. After making the proper serial connections, use the command *00X to read the default address. Use *00Xdd, where dd = the new address. If the old address is known, use DDXdd, where DD is the old address and dd is the new address. See the section on "Serial Communication: Commands" for more on command structure.

All commands listed in Appendix B are available for RS-485 use by inserting an asterisk and the unit's address in front of the command.

Ex) Set the units of measure on channel two (2) to %. *10UM203Cr

The command, *00XCr can be used to poll the unit for its current address setting using either RS-232 or RS-485.

10.5. SELECTING THE ANALOG SIGNAL LEVEL

The POWERPOD-400is capable of 0 to 5 V, 0 to 10 V or 4 to 20 mA operation. Each level is user selectable. The following diagram applies no matter which level of operation is chosen.

From

SETUP/CAL

Menu

10.5.1. Zero to Five Volt Operation

While in the SETUP/CAL menu, select ANALOG by pressing the number 4 key and the ENTER key. From the ANALOG SIGNAL menu, choose 0V – 5V by pressing the number 1 key and pressing the ENTER key. The normal display returns with 0 volts intended to correspond to 0% signal and 5 volts intended to correspond to the 100% signal. Calibration may be required.

10.5.2. Zero to Ten Volt Operation

While in the SETUP/CAL menu, select ANALOG by pressing the number 4 key and the ENTER key. From the ANALOG SIGNAL menu, choose 0V – 10V by pressing the number 2 key and pressing the ENTER key. The normal display returns with 0 volts intended to correspond to 0% signal and 10 volts intended to correspond to the 100% signal. Calibration may be required.

10.5.3. Four to Twenty Milliamp Operation

While in the SETUP/CAL menu, select ANLG by pressing the number 4 key and the ENTER key. From the ANALOG SIGNAL menu, choose 4mA – 20 mA operation by pressing the number 3 key and pressing ENTER. The normal display returns with 4 milliamps intended to correspond to 0% signal and 20 milliamps intended to correspond to the 100% signal. Calibration may be required.

164-122012 PowerPOD 400

Page 23 of 42

Image 23
Contents Teledyne Hastings Manual Print History Table of Contents Four to Twenty Milliamp Operation Quick Start Instructions PowerSafety Front Panel Lockout FeaturesPower Selection PowerPod-400 Specifications Table SpecificationsChannel Number Select Keys Front PanelRear Panel Communications WiringPower Transducer Connections AlarmsAnalog Signal Followers Override Open Power ON/OFFChanging the Command SET Point Manual OperationSetting a Channel to Display the Totalizer Function Setting a Channel to Auto ControlSetting a Channel to Directly Meter Incoming Signals Mode Reset Value Choosing RS-485 Communication External/Remote OperationChoosing RS-232 Communication Blanking a Display Line/Channel Setup/Cal ModeEntering the SETUP/CAL Mode Display SetupSelecting Gas ID Setting A/D Conversion Rate FilterSelecting Unit-of-Measure Count Up to a Set Point Totalizer SetupFront Panel Lock Out Count Down from a Set Point External Communication SetupRS-232 Settings RS-485 SettingsFour to Twenty Milliamp Operation Selecting the Analog Signal LevelZero to Five Volt Operation Zero to Ten Volt OperationSetting a Single Channel’s Hysteresis Setting Limit AlarmsSetting a Single Channel’s High Limit Setting a Single Channel’s Low LimitEnabling Ratio Control Setting Ratio Control ParametersResetting the Span Only Calibrating a Channel to ITS Incoming SignalsResetting the Zero Only Setting a Multiplication Factor Resetting Zero and SpanSet Point Queries CommandsAlarm/Flag Queries Serial CommunicationPowerPOD Appendix a PowerPod-400 Serial CommandsPowerPod-400 Serial Commands Contd Appendix a Notes Name Rate Total ApppendixCBGas ID Table Page 1 Gas ID Table Page 2 Appendix D Setting the Zero & Span PowerPod-400 Power Supply/TotalizerZeroing Unit 4 to 20 mA Range Spanning Unit Transducer POWERPOD-400 PowerPOD Drawings Warranty Repair Policy Warranty