Teledyne powerpod 400 instruction manual Mode Reset Value

Page 17

Example (See Graph):

A linear flow transducer with a DC output of 0 to 5 volts is calibrated for a maximum flow of 25 SLH. The transducer is connected to one channel of a POWERPOD-400which is CAL’d to read 25.000 at 5 volts input and has been programmed to display rate in SLH. The transducer is sending a constant 1.000- volt signal to the POWERPOD-400. What is the total flow after 40 minutes?

During one sampling interval, the TOTALIZER reads the one-volt signal and calculates the fraction of the possible full-scale value.

1 volt/5 volts or 0.20

The fractional signal value is multiplied times the CAL value 1/5 * 25 = 5

Then, because the channel’s programmed rate is in SLH (standard liters per hour), a factor of 1/3600 is applied to the result.

5 * 1/36000 = 1.389 x 10^-4 standard liters per sampling interval.

A constant signal received over a 40 minute time period would amount to 40 minutes * 60 seconds * 10 samples per second = 24000 samples. The TOTAL would be

2.400 E03 * 1.389 E-04 = 3 1/3 Std Liters.

If no rate is programmed on that channel, The TOTALIZER function is disabled for that channel and the display returns to its regular metering configuration.

To enable the TOTALIZER, press the MODE key. The MODE menu appears in the display. To select TOTAL, press 2, and then ENTER. The TOTALIZER menu allows the operator to select the desired channel. Press the number key that corresponds with the desired channel followed by ENTER. A TOTALIZER menu is presented that allows the operator to go directly to displaying whatever the Totalizer has currently stored as the total or to reset the counter. If DISPLAY is selected by pressing the number 1 key followed by ENTER, the display will return to normal with the most recently stored total being displayed on the appropriate line. If RESET is chosen by pressing the number 2 key followed by ENTER, the display returns with the total value reset according to the pre-selected counting mode presented in the table below with the Totalizer counting.

Mode

Reset Value

Count Down

Set Point

Count Up

Zero

Continuous

Zero

164-122012 PowerPOD 400

Page 17 of 42

Image 17
Contents Teledyne Hastings Manual Print History Table of Contents Four to Twenty Milliamp Operation Quick Start Instructions PowerSafety Front Panel Lockout FeaturesPower Selection PowerPod-400 Specifications Table SpecificationsChannel Number Select Keys Front PanelRear Panel Communications WiringPower Transducer Connections AlarmsAnalog Signal Followers Changing the Command SET Point Power ON/OFFManual Operation Override OpenSetting a Channel to Display the Totalizer Function Setting a Channel to Auto ControlSetting a Channel to Directly Meter Incoming Signals Mode Reset Value Choosing RS-485 Communication External/Remote OperationChoosing RS-232 Communication Entering the SETUP/CAL Mode Setup/Cal ModeDisplay Setup Blanking a Display Line/ChannelSelecting Gas ID Setting A/D Conversion Rate FilterSelecting Unit-of-Measure Count Up to a Set Point Totalizer SetupFront Panel Lock Out RS-232 Settings External Communication SetupRS-485 Settings Count Down from a Set PointZero to Five Volt Operation Selecting the Analog Signal LevelZero to Ten Volt Operation Four to Twenty Milliamp OperationSetting a Single Channel’s High Limit Setting Limit AlarmsSetting a Single Channel’s Low Limit Setting a Single Channel’s HysteresisEnabling Ratio Control Setting Ratio Control ParametersResetting the Span Only Calibrating a Channel to ITS Incoming SignalsResetting the Zero Only Setting a Multiplication Factor Resetting Zero and SpanAlarm/Flag Queries CommandsSerial Communication Set Point QueriesPowerPOD Appendix a PowerPod-400 Serial CommandsPowerPod-400 Serial Commands Contd Appendix a Notes Name Rate Total ApppendixCBGas ID Table Page 1 Gas ID Table Page 2 Appendix D Setting the Zero & Span PowerPod-400 Power Supply/TotalizerZeroing Unit 4 to 20 mA Range Spanning Unit Transducer POWERPOD-400 PowerPOD Drawings Warranty Repair Policy Warranty