Congratulations!

You have just calibrated your PowerPod-400to read zero at zero volts and the transducer’s max display value at the transducer’s max signal input.

The only thing left to do is to insure that your transducer is calibrated and that it is wired correctly to the PowerPod-400.

*Individual channels may be calibrated with calibrated transducers connected, breakout connectors (not supplied) and precision multi-meters with NIST traceable calibrations (also not supplied). Such procedures are not covered in this document.

!If the calibration for the channel just zeroed had been severely compromised prior to beginning this procedure, zero may not be immediately displayed and you may have to repeat the procedure after resetting the span. Check that the span has been set correctly!

#If a signal level within .1% of the maximum DC voltage level is not attainable, the power supply may need to be repaired.

164-122012 PowerPOD 400

Page 40 of 42

Image 40
Contents Teledyne Hastings Manual Print History Table of Contents Four to Twenty Milliamp Operation Power Quick Start InstructionsSafety Features Front Panel LockoutPower Selection Specifications PowerPod-400 Specifications TableFront Panel Channel Number Select KeysRear Panel Power WiringCommunications Alarms Transducer ConnectionsAnalog Signal Followers Power ON/OFF Changing the Command SET PointManual Operation Override OpenSetting a Channel to Directly Meter Incoming Signals Setting a Channel to Auto ControlSetting a Channel to Display the Totalizer Function Mode Reset Value Choosing RS-232 Communication External/Remote OperationChoosing RS-485 Communication Setup/Cal Mode Entering the SETUP/CAL ModeDisplay Setup Blanking a Display Line/ChannelSelecting Unit-of-Measure Setting A/D Conversion Rate FilterSelecting Gas ID Front Panel Lock Out Totalizer SetupCount Up to a Set Point External Communication Setup RS-232 SettingsRS-485 Settings Count Down from a Set PointSelecting the Analog Signal Level Zero to Five Volt OperationZero to Ten Volt Operation Four to Twenty Milliamp OperationSetting Limit Alarms Setting a Single Channel’s High LimitSetting a Single Channel’s Low Limit Setting a Single Channel’s HysteresisSetting Ratio Control Parameters Enabling Ratio ControlResetting the Zero Only Calibrating a Channel to ITS Incoming SignalsResetting the Span Only Resetting Zero and Span Setting a Multiplication FactorCommands Alarm/Flag QueriesSerial Communication Set Point QueriesPowerPOD PowerPod-400 Serial Commands Appendix aPowerPod-400 Serial Commands Contd Appendix a Notes ApppendixCB Name Rate TotalGas ID Table Page 1 Gas ID Table Page 2 Setting the Zero & Span PowerPod-400 Power Supply/Totalizer Appendix DZeroing Unit 4 to 20 mA Range Spanning Unit Transducer POWERPOD-400 PowerPOD Drawings Warranty Warranty Repair Policy