Teledyne powerpod 400 Appendix D, Setting the Zero & Span PowerPod-400 Power Supply/Totalizer

Page 36

14.0 Appendix D

Setting the Zero & Span

on the

PowerPod-400 Power Supply/Totalizer

All procedures outlined in this document must be performed with the power supply turned on and warmed up for at least one hour.

All procedures outlined in this document must be performed with NO transducers connected to the channel being adjusted*.

You CANNOT successfully zero a channel without, first, supplying zero volts or four milliamps to the meter input for that channel.

You CANNOT successfully span a channel without, first, supplying the meter input for that channel with its appropriate span voltage or span current.

14.1 Zeroing Unit (Voltage Range).

The following instructions explain how to short a given channel’s input signal to ground (0 volts DC) and set the display to read zero at this voltage.

Locate the 15-pin, ‘D’ type connector that corresponds to the channel to be zeroed. J1 = Channel 1

J2 = Channel 2

J3 = Channel 3

J4 = Channel 4

Short pin 5 (signal common) to pin 6 (signal in).

Press MODE

Press 3

Press ENTER

Press 1, 2, 3 or 4, depending on which channel is to be zeroed.

Press ENTER

Press 7

Press ENTER

Press 1

Press ENTER

Press ENTER again

You have just told a channel to display zero when the input signal is zero volts!.

164-122012 PowerPOD 400

Page 36 of 42

Image 36
Contents Teledyne Hastings Manual Print History Table of Contents Four to Twenty Milliamp Operation Power Quick Start InstructionsSafety Features Front Panel LockoutPower Selection Specifications PowerPod-400 Specifications TableFront Panel Channel Number Select KeysRear Panel Wiring PowerCommunications Alarms Transducer ConnectionsAnalog Signal Followers Power ON/OFF Changing the Command SET PointManual Operation Override OpenSetting a Channel to Auto Control Setting a Channel to Directly Meter Incoming SignalsSetting a Channel to Display the Totalizer Function Mode Reset Value External/Remote Operation Choosing RS-232 CommunicationChoosing RS-485 Communication Setup/Cal Mode Entering the SETUP/CAL ModeDisplay Setup Blanking a Display Line/ChannelSetting A/D Conversion Rate Filter Selecting Unit-of-MeasureSelecting Gas ID Totalizer Setup Front Panel Lock OutCount Up to a Set Point External Communication Setup RS-232 SettingsRS-485 Settings Count Down from a Set PointSelecting the Analog Signal Level Zero to Five Volt OperationZero to Ten Volt Operation Four to Twenty Milliamp OperationSetting Limit Alarms Setting a Single Channel’s High LimitSetting a Single Channel’s Low Limit Setting a Single Channel’s HysteresisSetting Ratio Control Parameters Enabling Ratio ControlCalibrating a Channel to ITS Incoming Signals Resetting the Zero OnlyResetting the Span Only Resetting Zero and Span Setting a Multiplication FactorCommands Alarm/Flag QueriesSerial Communication Set Point QueriesPowerPOD PowerPod-400 Serial Commands Appendix aPowerPod-400 Serial Commands Contd Appendix a Notes ApppendixCB Name Rate TotalGas ID Table Page 1 Gas ID Table Page 2 Setting the Zero & Span PowerPod-400 Power Supply/Totalizer Appendix DZeroing Unit 4 to 20 mA Range Spanning Unit Transducer POWERPOD-400 PowerPOD Drawings Warranty Warranty Repair Policy