Teledyne powerpod 400 instruction manual Transducer POWERPOD-400

Page 39

An example of one possible transducer/ POWERPOD-400combination.

 

TRANSDUCER

POWERPOD-400

Max signal out

Max value to be

Multiplier

Current Span Value

displayed

 

 

 

 

5 VDC

 

250.00

1.0000

100.00

Setting the THPS-400 Analog level to correspond with the transducer to be attached.

Press MODE

Press 3

Press ENTER

Press 1, 2, 3 or 4 depending on which channel is to be adjusted.

Press ENTER

Press 7

Press ENTER

Press 1 for 0 – 5 VDC, 2 for 0 – 10 VDC or 3 for 4 – 20 mADC

Press ENTER

In the example listed above, you would have selected option number 1 for the 0 – 5 volt analog signal level since the maximum signal from the transducer is 5 VDC.

You have just set the analog operating level (range) for the selected channel.

Setting the display to read the maximum transducer value at the maximum transducer input.

Locate the 15-pin, ‘D’ type connector that corresponds to the channel to be spanned. J1 = Channel 1

J2 = Channel 2

J3 = Channel 3

J4 = Channel 4

For units set to meter a voltage range, short pin 14 (command out) to pin 6 (signal in) of the selected channel and place a calibrated volt meter to read from pin 5 (common) to pin 6. For units set to read a 4 to 20 mA range, connect a milliamp meter between pins 14 and 6. Press the CHANNEL # key corresponding to the selected channel. An asterisk (*) appears. Use the number keys to enter THPS-400 current span value. Use the decimal!

Press ENTER

Adjust the command to achieve a value as close to the max analog signal level as possible (±0.01 volts for the 5 volt range, ±0.02 volts for the 10 volt range and ±0.004 ampsfor the milliamp range. Do this using the CHANNEL# key, the numeric key pad and the ENTER key until the desired signal is established#.

Press MODE Press 3 Press ENTER

Press 1, 2, 3 or 4 depending on which channel is to be adjusted. Press ENTER

Press 3

Press ENTER

Use the numeric keys to enter the new SPAN value (Max transducer value to be displayed). Press ENTER

Press ENTER

164-122012 PowerPOD 400

Page 39 of 42

Image 39
Contents Teledyne Hastings Manual Print History Table of Contents Four to Twenty Milliamp Operation Quick Start Instructions PowerSafety Front Panel Lockout FeaturesPower Selection PowerPod-400 Specifications Table SpecificationsChannel Number Select Keys Front PanelRear Panel Wiring PowerCommunications Transducer Connections AlarmsAnalog Signal Followers Override Open Power ON/OFFChanging the Command SET Point Manual OperationSetting a Channel to Auto Control Setting a Channel to Directly Meter Incoming SignalsSetting a Channel to Display the Totalizer Function Mode Reset Value External/Remote Operation Choosing RS-232 CommunicationChoosing RS-485 Communication Blanking a Display Line/Channel Setup/Cal ModeEntering the SETUP/CAL Mode Display SetupSetting A/D Conversion Rate Filter Selecting Unit-of-MeasureSelecting Gas ID Totalizer Setup Front Panel Lock OutCount Up to a Set Point Count Down from a Set Point External Communication SetupRS-232 Settings RS-485 SettingsFour to Twenty Milliamp Operation Selecting the Analog Signal LevelZero to Five Volt Operation Zero to Ten Volt OperationSetting a Single Channel’s Hysteresis Setting Limit AlarmsSetting a Single Channel’s High Limit Setting a Single Channel’s Low LimitEnabling Ratio Control Setting Ratio Control ParametersCalibrating a Channel to ITS Incoming Signals Resetting the Zero OnlyResetting the Span Only Setting a Multiplication Factor Resetting Zero and SpanSet Point Queries CommandsAlarm/Flag Queries Serial CommunicationPowerPOD Appendix a PowerPod-400 Serial CommandsPowerPod-400 Serial Commands Contd Appendix a Notes Name Rate Total ApppendixCBGas ID Table Page 1 Gas ID Table Page 2 Appendix D Setting the Zero & Span PowerPod-400 Power Supply/TotalizerZeroing Unit 4 to 20 mA Range Spanning Unit Transducer POWERPOD-400 PowerPOD Drawings Warranty Repair Policy Warranty