Teledyne powerpod 400 instruction manual Transducer Connections, Alarms

Page 13

7.3.TRANSDUCER CONNECTIONS

Connectors J1, 2, 3 and 4 (Item 7) are 15 pin D style connectors wired in the standard Hastings Instruments pin-out (H pin-out).

J1, J2, J3, J4

1

NC

 

 

2

NC

Valve Cntrl

 

 

 

Voltage

 

3

NC

mA Sig

 

4

NC

mA Sig

 

5

Sig. Com.

 

Transducer

6

Sig. In

 

7

Case Gnd.

 

Connectors

 

8

Cntrl Over-ride

1.5mA

(DB-15)

9

-15 VDC

 

(H-Pinout)

 

10

NC

 

 

 

 

11

+15 VDC

 

 

12

Valve Return

 

 

13

NC

Ext-In

 

14

Set Point Out

 

 

15

+5 VDC Ref.

Not Used.

7.4.ALARMS

Connector J8 (item 9) provides the user with open-collector, opto-isolated alarms for individual channels. Each channel is provided with one user settable “High” and one “Low” alarm.

J8

1

Chnl 1, High Alarm

 

2

Chnl 1, Low Alarm

 

3

Chnl 1, Alarm Return

 

4

Chnl 2, High Alarm

 

5

Chnl 2, Low Alarm

 

6

Chnl 2, Alarm Return

 

7

NC

 

8

NC

 

9

NC

 

10

NC

Open

11

NC

Collector,

12

NC

Opto-

13

NC

Isolated, High

14

NC

& Low

15

NC

Alarms

16

NC

(HD DB-26)

17

NC

 

18

NC

 

19

Chnl 3, High Alarm

 

20

Chnl 3, Low Alarm

 

21

Chnl 3, Alarm Return

 

22

Chnl 4, High Alarm

 

23

Chnl 4, Low Alarm

 

24

Chnl 4, Alarm Return

 

25

NC

 

26

NC

164-122012 PowerPOD 400

Page 13 of 42

Image 13
Contents Teledyne Hastings Manual Print History Table of Contents Four to Twenty Milliamp Operation Quick Start Instructions PowerSafety Front Panel Lockout FeaturesPower Selection PowerPod-400 Specifications Table SpecificationsChannel Number Select Keys Front PanelRear Panel Power WiringCommunications Transducer Connections AlarmsAnalog Signal Followers Changing the Command SET Point Power ON/OFFManual Operation Override OpenSetting a Channel to Directly Meter Incoming Signals Setting a Channel to Auto ControlSetting a Channel to Display the Totalizer Function Mode Reset Value Choosing RS-232 Communication External/Remote OperationChoosing RS-485 Communication Entering the SETUP/CAL Mode Setup/Cal ModeDisplay Setup Blanking a Display Line/ChannelSelecting Unit-of-Measure Setting A/D Conversion Rate FilterSelecting Gas ID Front Panel Lock Out Totalizer SetupCount Up to a Set Point RS-232 Settings External Communication SetupRS-485 Settings Count Down from a Set PointZero to Five Volt Operation Selecting the Analog Signal LevelZero to Ten Volt Operation Four to Twenty Milliamp OperationSetting a Single Channel’s High Limit Setting Limit AlarmsSetting a Single Channel’s Low Limit Setting a Single Channel’s HysteresisEnabling Ratio Control Setting Ratio Control ParametersResetting the Zero Only Calibrating a Channel to ITS Incoming SignalsResetting the Span Only Setting a Multiplication Factor Resetting Zero and SpanAlarm/Flag Queries CommandsSerial Communication Set Point QueriesPowerPOD Appendix a PowerPod-400 Serial CommandsPowerPod-400 Serial Commands Contd Appendix a Notes Name Rate Total ApppendixCBGas ID Table Page 1 Gas ID Table Page 2 Appendix D Setting the Zero & Span PowerPod-400 Power Supply/TotalizerZeroing Unit 4 to 20 mA Range Spanning Unit Transducer POWERPOD-400 PowerPOD Drawings Warranty Repair Policy Warranty