Page 16

8.5.SETTING A CHANNEL TO AUTO CONTROL

Press the desired CHANNEL # key. An asterisk appears in the first space on the line representing the selected channel. The meter display is immediately replaced with the current set point. Pressing the AUTO key causes pin number eight (8) of the 15 pin Sub-D connector to float and returns the previously programmed metering function to the display. Pin 14 signal levels are now available for control.

8.6.SETTING A CHANNEL TO DIRECTLY METER INCOMING SIGNALS

Press the MODE key. The MODE menu appears in the display. To select METER, press 1, and then ENTER. The METER menu allows the selection of the desired channel. Press the number key that corresponds with the desired channel followed by ENTER. The previously programmed display returns with the selected channel reading the signal between pins 5 and 6 of the corresponding channel’s 15 pin D-connector (J1 – J4).

8.7.SETTING A CHANNEL TO DISPLAY THE TOTALIZER FUNCTION

The POWERPOD-400 provides an integrated (Riemann Sum) value of the incoming signal for each channel.

 

T = Total

 

 

T = ( f S *C * f I ) , where:

fS

= Fractional Signal Factor

= CAL value or Span Value

 

C

 

 

 

f I

= Fractional Time Interval

Each channel samples the incoming signal at a rate corresponding to the preset A/D conversion rate in Hz. At each sampling interval, the TOTALIZER function multiplies the average signal, as a fraction of the full-scale value (5v, 10v or 20 mA), times the span, or cal, value. This results in a rate for that interval. Next, a time element factor is determined according to the flow rate programmed for that channel as follows.

For a time element of seconds, the factor is 1/10.

For minutes, the factor is 1/600.

Hours use a factor of 1/36000.

Each calculation is summed and stored for a TOTAL value.

164-122012 PowerPOD 400

Page 16 of 42

Image 16
Contents Teledyne Hastings Manual Print History Table of Contents Four to Twenty Milliamp Operation Power Quick Start InstructionsSafety Features Front Panel LockoutPower Selection Specifications PowerPod-400 Specifications TableFront Panel Channel Number Select KeysRear Panel Power WiringCommunications Alarms Transducer ConnectionsAnalog Signal Followers Power ON/OFF Changing the Command SET PointManual Operation Override OpenSetting a Channel to Directly Meter Incoming Signals Setting a Channel to Auto ControlSetting a Channel to Display the Totalizer Function Mode Reset Value Choosing RS-232 Communication External/Remote OperationChoosing RS-485 Communication Setup/Cal Mode Entering the SETUP/CAL ModeDisplay Setup Blanking a Display Line/ChannelSelecting Unit-of-Measure Setting A/D Conversion Rate FilterSelecting Gas ID Front Panel Lock Out Totalizer SetupCount Up to a Set Point External Communication Setup RS-232 SettingsRS-485 Settings Count Down from a Set PointSelecting the Analog Signal Level Zero to Five Volt OperationZero to Ten Volt Operation Four to Twenty Milliamp OperationSetting Limit Alarms Setting a Single Channel’s High LimitSetting a Single Channel’s Low Limit Setting a Single Channel’s HysteresisSetting Ratio Control Parameters Enabling Ratio ControlResetting the Zero Only Calibrating a Channel to ITS Incoming SignalsResetting the Span Only Resetting Zero and Span Setting a Multiplication FactorCommands Alarm/Flag QueriesSerial Communication Set Point QueriesPowerPOD PowerPod-400 Serial Commands Appendix aPowerPod-400 Serial Commands Contd Appendix a Notes ApppendixCB Name Rate TotalGas ID Table Page 1 Gas ID Table Page 2 Setting the Zero & Span PowerPod-400 Power Supply/Totalizer Appendix DZeroing Unit 4 to 20 mA Range Spanning Unit Transducer POWERPOD-400 PowerPOD Drawings Warranty Warranty Repair Policy