14.3 Spanning Unit.

The following instructions explain how to use the POWERPOD-400 command signal to supply a given channel’s input with the proper span voltage and to set the display to read a transducer’s span value.

You must know four things prior to setting the display’s span value:

1.You must know the maximum signal level for the specific transducer connected to the channel to be spanned (5 VDC, 10VDC or 20mA).

2.You must know the maximum value to be displayed when the maximum signal level for the transducer is connected to the channel to be spanned.

3.You must insure that the channel’s multiplier is set to 1.0000.

4.You must know the channel’s ‘Span Value’. That is, the value that would NOW be displayed (prior to any adjustment) if the maximum signal was fed into the input of the channel to be adjusted.

1 and 2. If this information is not on the transducer itself, consult the manufacturer of the transducer. Their literature or other documentation should specify the maximum output signal from the transducer and the maximum value to be displayed at that signal level.

3.Setting the MULTIPLIER to 1.0000. Press MODE

Press 3

Press ENTER

Press 1, 2, 3 or 4 depending on which channel is to be adjusted. Press ENTER

Press 7

Press ENTER Press 4 Press ENTER

Press 1.0000 using the numeric key pad. Press ENTER

You have just set a channel’s multiplier to 1.0000.

4.If you do not know what the selected channel’s maximum display (Span) value is currently set to, follow these steps.

Notice that the last step in this exercise is to press ESC, not ENTER.

Press MODE

Press 3

Press ENTER

Press 1, 2, 3 or 4 depending on which channel is to be adjusted.

Press ENTER

Press 7

Press ENTER

Press 3

Press ENTER

Read the VALUE from the display.

Press ESC

164-122012 PowerPOD 400

Page 38 of 42

Image 38
Contents Teledyne Hastings Manual Print History Table of Contents Four to Twenty Milliamp Operation Power Quick Start InstructionsSafety Features Front Panel LockoutPower Selection Specifications PowerPod-400 Specifications TableFront Panel Channel Number Select KeysRear Panel Communications WiringPower Alarms Transducer ConnectionsAnalog Signal Followers Manual Operation Power ON/OFFChanging the Command SET Point Override OpenSetting a Channel to Display the Totalizer Function Setting a Channel to Auto ControlSetting a Channel to Directly Meter Incoming Signals Mode Reset Value Choosing RS-485 Communication External/Remote OperationChoosing RS-232 Communication Display Setup Setup/Cal ModeEntering the SETUP/CAL Mode Blanking a Display Line/ChannelSelecting Gas ID Setting A/D Conversion Rate FilterSelecting Unit-of-Measure Count Up to a Set Point Totalizer SetupFront Panel Lock Out RS-485 Settings External Communication SetupRS-232 Settings Count Down from a Set PointZero to Ten Volt Operation Selecting the Analog Signal LevelZero to Five Volt Operation Four to Twenty Milliamp OperationSetting a Single Channel’s Low Limit Setting Limit AlarmsSetting a Single Channel’s High Limit Setting a Single Channel’s HysteresisSetting Ratio Control Parameters Enabling Ratio ControlResetting the Span Only Calibrating a Channel to ITS Incoming SignalsResetting the Zero Only Resetting Zero and Span Setting a Multiplication FactorSerial Communication CommandsAlarm/Flag Queries Set Point QueriesPowerPOD PowerPod-400 Serial Commands Appendix aPowerPod-400 Serial Commands Contd Appendix a Notes ApppendixCB Name Rate TotalGas ID Table Page 1 Gas ID Table Page 2 Setting the Zero & Span PowerPod-400 Power Supply/Totalizer Appendix DZeroing Unit 4 to 20 mA Range Spanning Unit Transducer POWERPOD-400 PowerPOD Drawings Warranty Warranty Repair Policy