Emerson 00825-0100-4530 Rev EC manual Chinese certifications, March, I21 Intrinsically safe

Page 29
I21 Intrinsically safe:

March 2013

Quick Installation Guide

I21 Intrinsically safe:

4-20 mA/HART model: Ex ia IIC T4 Ga

Ex ta IIIC T79 °C

-50 °C < Ta < +70 °C

Ui: 30 V, Ii: 130 mA, Pi: 1.0 W, Li: 0 μH, Ci: 7.26 nF FOUNDATION fieldbus model:

Ex ia IIC T4 Ga Ex ta IIIC T69 °C

-50 °C < Ta < +60 °C

Ui: 30 V, Ii: 300 mA, Pi: 1.5 W, Li: 0 μH, Ci: 4.95 nF Installation Drawing: 9240030-938

IB1 FISCO model:

FISCO Field Device Ex ia IIC T4 Ga

Ex ia/ib IIC T4 Ga/Gb Ex ta IIIC T69 °C -50 °C < Ta < +60 °C

Ui: 17.5 V, Ii: 380 mA, Pi: 5.32 W, Li: <1 μH, Ci: 4.95 nF Installation Drawing: 9240030-938

Chinese certifications

National Supervision and Inspection Center for Explosion Protection and Safety of Instrumentation (NEPSI) approvals

Special conditions for safe use (X):

Refer to Certificates GYJ 111230X and GYJ 081131X.

E31 Flameproof:

HART model:

Ex d ia IIC T4 (-40 °C < Ta < +70 °C) DIP A20 TA79 °C

FOUNDATION fieldbus model:

Ex d ia IIC T4 (-40 °C < Ta < +60 °C) DIP A20 TA69 °C

Approval valid for HART, FOUNDATION fieldbus, and Modbus options.

I31 Intrinsically safe:

HART model:

Ex ia IIC T4 (-50 °C < Ta < +70 °C) DIP A20 TA79 °C

4-20 mA/HART model: Ui = 30 V, Ii = 130 mA, Pi = 1.0 W, Ci = 7.26 nF, Li = 0 μH FOUNDATION fieldbus model:

Ex ia IIC T4 (-50 °C < Ta < +60 °C) DIP A20 TA69 °C

Ui = 30 V, Ii = 300 mA, Pi = 1.5 W, Ci = 4.95 nF, Li = 0 μH Approval valid for HART and FOUNDATION fieldbus options.

1)Ordering information code for Product Certifications, see the Rosemount 5300 Series Product Data Sheet (Document No. 00813-0100-4530) or the Rosemount 5300 Series Reference Manual

(Document No. 00809-0100-4530).

29

Image 29
Contents Quick Installation Guide 00825-0100-4530,Rev EC MarchSuperior Performance Guided Wave Radar Level and Interface TransmitterProbes with non-conductingsurfaces About this guideQuick Installation Guide March ContentsStep 1: Confirm system readiness Switch HART revision modeConfirm HART revision capability Confirm correct Device DriverTank connection with flange Step 2 Mount the transmitter head/probeThreaded tank connection Tri-Clamptank connectionBracket mounting, on pipe Step 2 continuedBracket mounting, on wall Step 2 continued Remote housingGrounding Step 3 Connect the wiringTerminal block To connect the transmitterIntrinsically safe power supply Non-intrinsicallysafe power supplyHART communication Step 3 continuedQuick Installation Guide Transient protection terminal blockPower PowerQuick Installation Guide Load limitationsNon-HazardousInstallations InstallationsIntrinsically safe power supply Step 3 continued FOUNDATION fieldbusNon-intrinsicallysafe power supply Power Consumption 0.5 W with HART address = RS-485with Modbus communicationQuick Installation Guide Step 3 continuedInstalling the Rosemount Radar Master software Step 4 Configure1Only for HART communication Run WizardFOUNDATION fieldbus parameters FOUNDATION fieldbus parameters FOUNDATION fieldbus parameters 3.Click Modbus Setup Setup - Modbus communication parametersConfiguration Safety Instrumented Systems 4-20mA onlyInstallation Table 3. Alarm levels and operation values Alarm and saturation levelsDamping Write protectionProof test Operation and maintenanceSite acceptance Inspection SpecificationsReferences Visual inspectionProduct certifications Safety Instrumented Systems SISEU conformity North-AmericancertificationsQuick Installation Guide Canadian Standards Association CSA approvalsDual seal maintenance I5, IE1 Intrinsically safe and FISCO modelATEX approvals European certificationsMarch Special conditions for safe useINMETRO approvals Brazilian certificationsQuick Installation Guide Special conditions for safe useMarch Chinese certificationsI21 Intrinsically safe IB1 FISCO modelQuick Installation Guide Japanese certificationsIC1 FOUNDATION fieldbus FISCO model N3 Type nIECEx approvals IECEx certificationsMarch Special conditions for safe useOverfill protection Other certificationsQuick Installation Guide Suitability for intended useFigure 1. EC Declaration of Conformity - page EC Declaration of ConformityQuick Installation Guide MarchMarch Figure 2. EC Declaration of Conformity - pageQuick Installation Guide March Figure 3. EC Declaration of Conformity - pageQuick Installation Guide 00825-0100-4530 Quick Installation Guide00825-0100-4530,Rev EC March Emerson Process Management Rosemount Measurement