Eye-Diagram Analyzer Reference
Calibrating the Eye-Diagram Analyzer
1Connect a low-loss cable between the front-panel CALIBRATOR OUTPUT and 
channel 1 connectors. 
2To calibrate the IF, press: 
MENU, page 1 of 2, Calib, IF calib, CAL CH1 
The calibration routine takes about 30 minut es to complete. 
3Disconnect the cable from channel 1, and connect it to the channel 2 
connector. 
4Repeat the above procedure for channel 2.