Eye-Diagram Analyzer Reference
Agilent 70820A Calibration
Note Warm up time Allow the microwave transition analyzer module to warm up for
at least one hour before performing a calibration.
To Calibrate the
IF
1Connect a low-loss cable between the front-panel CALIBRATOR OUTPUT and
channel 1 connectors.
2To calibrate the IF, press:
MENU, page 1 of 2, Calib, IF calib, CAL CH1, CAL CH1
The calibration routine takes about 30 minut es to complete.
3Disconnect the cable from channel 1, and connect it to the channel 2
connector.
4Repeat the above procedure for channel 2.
To View the Last
Calibration Date
1Turn any displayed traces off to make viewing the calibration date easi er.
2Continue by pressing:
MENU, page 1 of 2, Calib, IF calib, cal debug, show
3To view when channel 1 was last calibrated, press:
CH1 CAL INFO
4To view when channel 2 was last calibrated, press:
CH2 CAL INFO