Eye-Diagram Analyzer Reference
Displaying Traces
To Position and 
Scale Each Trace  The selected trace can be positioned vertically on the display. This is accom-
plished using the REF LEV|POS softkey to enter both the reference position and 
reference level. The reference position (POS) is entered as a percentage of the verti-
cal display. The top of the screen is 100%. The bottom of the screen is 0%. The ref-
erence level (REF LEV) is the amplitude value at the reference position. 
To scale and position a trace, press: 
Traces, AUTO-SCALE to automatically scale the selected trace.  
or
SCALE to manually set the scale.
Use the front-panel knob or numeric keypad to change the displayed ve rtical 
scaling.  
REF LEV|POS POS to move the trace’s reference position. 
Rotate the front-panel knob to change the position.  
REF LEV|POS LEV to change the reference level at the trace’s reference 
position. 
Rotate the front-panel knob to change the level.  
To Average Trace 
1  The Traces menu’s average softkey presents a menu used for controlling trace 
averaging. Trace averaging keeps a running average of the measurement val-
ues on each sweep. You can set the averaging count; its default value is 16. 
Trace averaging is only applied to trace 1.