2-27
Application Tutorials
Tutorial 8: Display the Data Pattern
Tutorial 8: Display the Data Pattern   The eye-diagram analyzer can display the data pattern. Th i s is  done by select-
ing the pattern mode and triggering the display trace update on the patte rn  
trigger. This insures the trace on the screen remains the same from sweep to 
sweep. In this mode, trace averaging can be used for more repeatable mea-
surements on noisy signals. 
Masks and limit lines can be used with the pattern mode and are useful for 
testing specific portions of the data sequence for  mask or template violations. 
This can uncover violations that happen only when a specific pattern of ones 
and zeros occur. The compatible mode is Pattern.
Select Pattern Mode 
1Press INSTR PRESET. 
2If you are using an  70841A/B pattern generator, connect the TRIGGER OUT 
signal to the  70820A module’s RF INPUT 2 connector. 
This procedure uses the eye-diagram analyzer’s pattern mode. In pattern 
mode, the trigger signal must come from the TRIGGER OUT and not the 
CLOCK OUT connector. The pattern mode only works with a pattern trigger, 
and not with a clock signal. The pattern trigger is derived from the clock signal 
divided by the pattern length. 
3Select the pattern mode by pressing: 
Setup, diagram, PATTERN 
Notice the CH2 is: softkey label has changed to indicate that the trigger source 
is connected to the TRIGGER OUT signal. 
4Select the bit interval and the delay by pressing: 
BIT INTVL and enter 20, Traces 
5Turn autoscaling on by pressing: 
AUTO-SCALE 
The display should look like the following figure.