Eye-Diagram Analyzer Reference
Agilent 70820A User-Corrections
To Delete a 
Channel’s User-
Corrections 
1Delete a channel’s user-corrections by pressing: 
MENU, page 1 of 2, Calib, user corr  
2Select the desired channel by pressing: 
CH1|CH2 
3Delete the selected channels user-corrections by pressing : 
other, DELETE ALL  
4To delete all corrections, press: 
CONFIRM DEL ALL  
5To cancel the deletion process, press: 
ABORT DEL ALL  
Entering and Editing User-Corrections The User Correction menu provides softkeys for entering and editing lines of 
user-correction data. To view user-correction data graphically at the same 
time it is created, refer to “Viewing User-Correction Data” on page 3-56. 
The SEL|EDT softkey determines whether the front-panel knob and arrow 
keys select a field for editing the current f ield’s data. New lines can be entered 
between the first and last lines. Refer to "To enter a line of data" in this sec-
tion. 
The microwave transition analyzer interpolates the data between adjacent 
points. Interpolation between the defined points can be either  slope or flat 
line. If slope is selected, the slope LOG|LIN softkey determines whether 
slopes are linear or logarithmic. To locate this softkey, press other in the User 
Correction menu.