Eye-Diagram Analyzer Reference
Saving to Mass Storage
6Issue the following Instrument BASIC command to copy the files from the RAM
disk to the new card:
COPY "*:MEMORY,0,0" TO ":EXTERNAL,904"
To Catalog all
Files Use this procedure to catalog default memory. To select default memory, refer
to “Selecting and Formatting Memory” on page 3-34.
1If you plan to catalog the files on a memory card, insert the card in the front-
panel card slot.
2Display a catalog of the files by pressing:
MENU, page 1 of 2, States, more 1 of 2, mass storage
Remember to insert the card before cataloging a memory card. If the message
"6218 catalog open failed" is displayed, the card is missing. The HP-
MSIB address for the card is the same address as the display’s GPIB address
and is normally set to 4. If the address is not correct, enter the correct address
using the numeric keypad.
To Save a File Use this procedure to save a file in the default memory. To select default mem-
ory, refer to “Selecting and Formatting Memory” on page 3-34.
1Display the Mass Storage menu by pressing:
MENU, page 1 of 2, States, more 1 of 2, mass storage, save
2Perform one of the following:
•If you are saving an instrument state, user-cor rection, or histogram, press the
appropriate softkey.
•If you are saving a trace, press save trace. Select the trace and then use the
VEC|FMT softkey to determine if the trace data is saved as complex data
(VEC) or scalar data (FMT). Press save.
•If you are saving a channel calibration, mask, limit line, us er -defined key, all
recall registers, press save misc and then the appropriate softkey.
3Enter the desired file name.
4To save the file, press:
ENTER LINE
An HP-IL keyboard is recommended to simplify the task of entering alphanu-
meric information. This keyboard plugs into the jack provided o n the display’s
front panel. Order the Agilent 46021A keyboard.