Exatron 3000B
www.exatron.com 6-9 Chapter 6 Display Dictionary
FULL TUBE COUNT
Press the [7] key. Display will read, FULL TUBE = _____. (The current full tube count
setting will be shown.)
Enter desired full tube count changes using the controller keypad number keys. The
handler allows changes to each digit from lowest to highest in order. First enter Ones, then
Tens, then Hundreds, then Thousands. For example: to enter a tube full count of 25, press
5, then 2, then 0, then 0. The display will read: 0025.
Press the [Enter] button to return to HANDLER PAUSED message.
Press the [RUN] button on the handler controller keypad to exit Pause
Mode and return to previous loop.
HANDLER PORTPlease refer to PICK INTERFACE? message later in this section.
INPUT 12345678One or more of the numbers shown above is displayed following the word INPUT. Please refer to the
Diagnostics/OCTO/INDEX CHECK: Input Sensor Check section in Chapter 5 of this manual for details.
INSERTION ERRORThe handler has detected a continuity error when trying to insert the device under test into the contacts.
To resolve this condition:
1. If you are using a plunge-to-board changeover kit, check the plunge depth setting. An
incorrect plunge depth setting is the most common reason for this condition. Please refer to
the "Automatic Z Adjust Feature" in Chapter 5 for details.
2. Inspect your P I contactors and clean the contact pins if necessary. Please refer to the
"Particle Interconnect Contact Maintenance" section in Chapter 7 for details.
3. Inspect your device for bent leads.
INTERFACE CHECKPlease refer to the Diagnostics/INTERFACE CHECK section in Chapter 5 of this manual for details.
INTO SHUT JAMA device which has exited the test site has jammed as it arrived at the top of the output shuttle. Open the
output door and visually inspect the shuttle area. Verify that the shuttle is fully seated in the handler.
Securing the shuttle onto the shuttle car by pressing it down with finger pressure until it snaps into place
often resolves this error condition. Close the output door and the handler will resume operation.
Alternatively, the shuttle may have become slightly out of alignment with the output track. Press the [0]
key on the handler controller keypad to turn ALL MOTORS OFF. Press the [Enter] button to display
ALIGN MOTORS ? Align the shuttle by:
Pressing the [7] key to jog the shuttle one step left.
Pressing the [9] key to jog the shuttle one step right.
When alignment is complete, press the [RUN] button to resume operation.