www.exatron.com 5-16 Chapter 5 Diagnostics
;
;;
; ALL MOTOR CHECK: Octoloader Motor Check
__________________________________________________________________
This diagnostic tests input port C6, output port C6, C9 located on PCB #5000-397, PAL 397-5, U9.
TO ENTER THIS CHECK:
WARNING: VERIFY THAT THE OCTOLOADER IS EMPTY OF ALL DEVICES BEFORE USING THIS
DIAGNOSTIC.
Handler displays DIAGNOSTICS
Press the [4] key.
Handler displays ALL MOTOR CHECK
Press the [3] key.
Handler displays OCTO MOTOR CHK
Press the [1] key: 1 LEFT STEP. The Octoloader will move one tube to the left.
Press the [2] key: the handler will display the current motor settings in RAM. This display is helpful when
trying to increase the motor's speed.
Press the [3] key: 1 RIGHT STEP. The Octoloader will move one tube to the right.
Press the [4] key: 2 LEFT STEPS. The Octoloader will move two tubes to the left.
Press the [6] key: 2 RIGHT STEPS. The Octoloader will move two tubes to the right.
Press the [7] key: JOG LEFT. The Octoloader will move one step to the left, about .006".
Press the [9] key: JOG RIGHT. The Octoloader will move one step to the right, about .006".
Press the [0] key: MOTOR OFF. The Octoloader Motor is now off.
Pressing the [Clear] button will flash the terminal's displayed message to confirm the handler is still in this
specific diagnostic check.
Press the [Enter] button to EXIT this diagnostic check.